Infrared K-Thermometer Model CA876
7
CHAPTER 3
SPECIFICATIONS
3.1 Environmental Specifications
3.1.1 INFRARED
Temperature Scale:
Celsius (
°
C) or Fahrenheit (
°
F) user-selectable
Temperature Range:
-4°F to 1022°F (-20°C to 550°C)
Display Resolution:
1°F / 1°C
Accuracy:
< 212°F (100°C) ± 10°F (5°C)
>
212°F (100°C) ± 2% of Reading or ± 6°F (3°C), whichever is greater @
64.4 to 82.4°F (18 to 28°C) ambient operating temperature.
Temperature Coefficient:
Changes in accuracy operating temperature above 82.4°F/28°C or below
64.4°F/18°C: ±0.2% of Reading or ± 0.36°F/0.2°C, whichever is greater.
Response Time:
1 second
Laser:
Red, <1mW (670nm) Class II, 2 - 50 ft range
Spectral Response:
6 to 14
μ
m nominal
Emissivity:
Pre-set @ 0.95, user-selectable from 0.10 to 1.00
Detection Element:
Thermopile
Optical Lens:
Fresnel Lens