Aegis EL-RG-49H Скачать руководство пользователя страница 7

24

5

Check refrigerator settings when restarting the refrigerator after power failure or the 

power is shut off. Changes of settings may damage the articles stored.
The refrigerator shall be reconnected after more than 5 minutes once it is shut off, to 

prevent damage to compressor or system.
Wear gloves when performing maintenance to prevent personal injuries due to sharp 

edges or corners.
Hold the knob when closing the door to prevent finger pinching.

The angle of inclination shall not be greater than 45°when handling the refrigerator.
Be careful not to be tripped up by the refrigerator during handling, to prevent 

refrigerator damage or personal injuries.
Do not lift or handle equipment with door knob to prevent refrigerator damage or 

personal injuries.

Do not use electrical appliances inside the food storage compartments of the appliance, 

unless they are of the type recommended by the manufacturer.

Do not damage refrigerating circuit.

The appliance must be positioned so that the plug is accessible.
The appliance must be placed on a solid and flat surface, or excessive vibration and 

noise may be produced when the appliance in operation. 
The appliance can be used by children aged from 8 years and above and persons with 

reduced physical, sensory or mental capacibilities or lack of experience and knowledge

if they have been given supervision or instruction concerning use of the appliance in a 

safe way and understand the hazards involved. Children shall not play with the appliance. 

Cleaning and user maintenance shall not be made by children without supervision.
If the supply cord is damaged, it must be replaced by the manufacturer, its service agent 

or similarly qualified persons in order to avoid a hazard. 
Keep ventilation openings, in the appliance enclosure or in the built-in structure, clear of 

obstruction.
If your cabinet is to be discard, you must remove the door and leave the shelves in 

place. This will reduce the posibility of danger to children. And the flammable foaming 

needs to be disposed by professional persons.
CP/IP foaming materials are flammable, need professional processing.

Refrigeration Schematic & Circuit Diagram

Schematic diagram

Evaporimeter

Muffler

Dry filter

Capillary tube

Condenser

Compressor

Exhaust 

pipe

Содержание EL-RG-49H

Страница 1: ...or may vary Translation of the original instruction The scope of the Product Application Pharmaceuticals and other related products which need a storage environment of 2 8 Certi cate of Quality Pharma...

Страница 2: ...harmaceutical Refrigerator Type Designation Trademark EL RG 49H Manufactures Name Aegis Scientific Inc Address Tel 1 Ivybrook Blvd Unit 145R Ivyland PA 18974 USA 1 800 796 2344 Fax 732 879 0248 Refer...

Страница 3: ...essor and main components and better refrigeration performance User friendly design computer control smart and carefree adjust ment not required High performance thermal insulation layer with excellen...

Страница 4: ...the appliance that you have just purchased due to technical improvements 1 Product Feature Content Safety Precautions Precautions for Use Product Installation Component Names Control Panel Applicatio...

Страница 5: ...ug in or pull out the power plug of refrigerator to prevent explosion and fire The refrigerator shall be installed by professional technical staffs or after sale maintenance staffs to prevent electric...

Страница 6: ...ay pose danger to human health or environment Pull out the power plug if the refrigerator has been out of service for long time to prevent electric shock leakage or fire caused by aging power lines If...

Страница 7: ...ust be placed on a solid and flat surface or excessive vibration and noise may be produced when the appliance in operation The appliance can be used by children aged from 8 years and above and persons...

Страница 8: ...hazardous substances can leak into the groundwater and get into the food chain damaging your health and well being When replacing old appliances with new ones the retailer is legally obligated to tak...

Страница 9: ...ion site Do not ground the refrigerator through gas pipes water supply pipes telephone lines or lightning conductors as electric shock may be caused The power plug can be accessed after installation t...

Страница 10: ...ownward upward Clean the back evaporating box Evaporating box is installed on the back of the refrigerator Remove 2 screws on the bottom loosen the upper 2 screws no need to remove them lift the evapo...

Страница 11: ...m is under normal condition When chargeable battery is closed or with too low a power press once the alarm testing key E6 is expected to flash 3 times on the display window at the simultaneously The r...

Страница 12: ...220V 110V failed under any of the above conditions remote alarm shall be started Coupled with 1Hz continuous beeping Take corres ponding measures according to item 1 2 and 3 Remote alarm startup when...

Страница 13: ...upper or lower temperature displayed ALH Alarm Limit High settable by the user high temperature alarming is triggered while when any of upper or lower temperature shown ALL Aalrm LImit Low settable by...

Страница 14: ...inside temperature display area flashes Date press Minus button or Plus button to adjust the current date and the press Set button to save and confirm Thereafter the inside temperature display area sh...

Страница 15: ...e display area steadily shows 2P press Set button the inside temperature display area flashes Month factory default 01 press Minus button or Plus button select Month from 01 12 recycling display after...

Страница 16: ...arameter value After setting press Set to go back and choose other parameters for the same token With all values set press Set button 5s exit and return to the normal display mode Temperature Display...