Advanced Instruments, Inc.
2. Quality Control Certification
Date:
Customer: Order No.:
Pass
Model:
GPR-2500A Oxygen Transmitter
Sensor:
( ) GPR-11-32-4 Oxygen Sensor
( ) XLT-11-24-4 oxygen Sensor
Serial Nos.:
Transmitter ______________________ Sensor
______________________
Accessories:
Owner’s Manual
Configuration:
A-1161-AIS-2 Rev C4 PCB Assembly Main Processing
A-1166-AIS-2 Rev H PCB Assembly Alarms/Power Connection
Software rev:
Ranges: 0-1%, 0-5%, 0-10%, 0-25%
Power: 12-24V DC
Barometric pressure and temperature compensation
NEMA 4X rated wall mount enclosure
Test:
Pass
Pass
Set default zero
Set default span @ 40uA
Alarm delay
Zero calibration
Alarm bypass
Span Calibration
Alarm configurations; ALARM 1,
ALARM 2
Analog signal output 4-20mA full scale
Alarm function; ALARM 1, ALARM 2
Calibrates with adequate span adjustment within
10-50% FS
Alarm relays; ALARM 1, ALARM 2
Baseline drift on zero gas <
±
2% FS over 24 hour
period
Alarm system fail, dry contact
Noise level <
±
1.0% FS
Overall inspection for physical
defects
Options:
Notes:
4