Advanced Instruments, Inc.
Allow approximately 5-10 min for oxygen to stabilize and the press ENTER to complete span gas
calibration
Both the Zero Calibrate and Span Calibrate functions result in the following displays showing either
'PASSED CALIBRATION" OR "FAILED CALIBRATION"
In either case, the transmitter returns to the SAMPLING mode and resume sampling.
Disconnect the span gas line and replace it with sample gas line.
Wait 10-15 minutes to ensure the reading is stable and then proceed to sampling.
20.01%
SPAN
CALIBRATION IN PROGRESS
ENTER TO CALIBRATE
MENU TO ABORT
PASSED
CALIBRATION
FAILED
CALIBRATION