Cannot perform Zero
calibration
Excessive back pressure on sensor
caused by high sample flow
Obstruction on sample vent line
Abnormality in span gas
Leak in sample/zero gas line
Inadequate zero gas flow
Contaminated zero gas
Attempting to Zero calibrate
prematurely
Set sample flow between 1-2 SCFH
Vent sample to atmosphere
If venting sample to a vent pipe with
pressure above atmosphere, consult
factory for recommendations.
Qualify span gas with a secondary
analyzer
Check for any leakage in sample/zero
gas line
Check the integrity of zero gas
Set the gas flow rate between 1-2
SCFH
Allow sufficient time for the oxygen
reading to stabilize before attempting
Zero calibration.