Advanced Instruments Inc.
2. Quality Control Certification
Date:
Customer: Order No.:
Pass
Model
GPR-1200 Portable PPM Oxygen Analyzer S/N ____________________
Sensor
( ) GPR-12-333 PPM Oxygen Sensor
( ) XLT-12-333 PPM Oxygen Sensor S/N ____________________
Accessories
Owner’s Manual
( ) PWRS-1002 9VDC Battery Charger/Adapter 110VAC
( ) PWRS-1003 9VDC Battery Charger/Adapter 220VAC
( ) PWRS-1008 9VDC Battery Charger/Adapter 12VDC Auto Cigarette Lighter
CONN-1034 Plug Mini Phone .141 dia. Black Handle
TOOL-1001 5/16 Combination Wrench
Configuration
A-1161-B Rev C3 PCB
Software Version
A-3878 Battery Assembly
B-3346 Sample Pump
Range: 0-10 ppm, 0-100 ppm, 0-1000 ppm, 0-25%
Wetted parts: stainless steel
Electronics Test
LED indicators: Low battery, Battery charge
Electronic offset set
Analog signal output 0-1V Full scale
Gas Phase Test
Recovery from air to < 10 PPM in < 1 hour
Baseline drift on zero gas <
±
2% FS over 24 hour period on 0-10 PPM range
Noise level <
±
0.5% FS
Span adjustment within 10-50% FS
Final
Overall inspection for physical defects
Options
Notes
4