Advanced Instruments Inc.
MAIN MENU
AUTO SAMPLE
MANUAL SAMPLE
CALIBRATION
>>>
CALIBRATION
SPAN CALIBRATE
ZERO CALIBRATE
DEFAULT SPAN
DEFAULT ZERO
OUTPUT SPAN
OUTPUT ZERO
3.
Advance the reverse shade cursor using the ARROW keys to highlight DEFAULT ZERO.
4.
Press the ENTER key to select the highlighted menu option.
The following display appears and after 3 seconds the system returns to the SAMPLING mode:
FACTORY
DEFAULTS
SET
0.5 PPM
AUTO SAMPLING
10 PPM RANGE
76 F 100 KPA
Analog Output with Zero O
2
In rare instances the 0-1 V signal output may not agree with the reading displayed on the LCD. This feature enables
the user to adjust the 0V signal output when the LCD displays 00.00. Note: Adjust the 1V signal output with the
OUTPUT SPAN option described below.
1.
Access the MAIN MENU by pressing the MENU key.
2.
Advance the reverse shade cursor using the ARROW keys to highlight CALIBRATION.
3.
Press the ENTER key to select the highlighted menu option and the following displays appear:
MAIN MENU
AUTO SAMPLE
MANUAL SAMPLE
CALIBRATION
>>>
CALIBRATION
SPAN CALIBRATE
ZERO CALIBRATE
DEFAULT SPAN
DEFAULT ZERO
OUTPUT SPAN
OUTPUT ZERO
4.
Advance the reverse shade cursor using the ARROW keys to highlight DEFAULT ZERO.
5.
Press the ENTER key to select the highlighted menu option and the following display appears: