Advance acoustic Mupid-2plus Скачать руководство пользователя страница 9

9

8

3 Electrophoresis

Indication line

(1)  Place the prepared gel on the gel bed of the migration 

tank together with the tray. Since the side near the 

power supply is cathode, wells should be placed on 

the right for migration of nucleic acids in most cases. 

Wipe off any excess gel that is adhering around and 

bottom surface of the gel tray.

(2)  Pour buffer solution to a level of about 3 mm above 

the surface of the gel. However, do not fill above the 

level of the buffer indication line. 250-300 mL of 

buffer is required.

(3)  Connect the power supply to the migration tank and 

inset the electric plug into an outlet. The electric 

current cannot be turned on with the lid open.

(4)  Apply the sample to the wells using a pipette. The 

amount of applicable sample depends on the 

thickness of the gel; for example, a 4 mm well will 

take about 5-8 

µ

l and a 6mm well about 8-12 

µ

l. 

Note that the wells do not correspond to the pitch of 

a multipipette.

(1)

(2)

(5)  After verifying that the operation switch of the power 

supply is OFF, close the migration tank lid. 

Do this as follows:

1) First insert the 2 tabs on the cover into the holes in 

the migration tank.

2) Then rotate the cover forward to close it.

(6) Set the desired voltage using the output selection 

switch.

(7)  Push the operation switch to start migration. At this 

time verify that the output indicator LED is lit and that 

bubbles are forming on the platinum electrodes.

(8)  Migration distance may be estimated using the line 

printed on the gel tray or the slits on the migration tank 

lid. Be sure to turn the operation switch OFF when 

sufficient migration has been attained.

(9)  Then proceed quickly to the operation of staining the gel.

9P

8P

ON

OFF

LED

Migration may be affected if the migration tank 
is not placed on a level surface.

When changing the output, make 
certain that the operation switch is OFF. 
In this case the output indicator LED 
will be unlit.

Take care not to put any foreign objects 
into or touch inside of the migration 
tank during migration.

Содержание Mupid-2plus

Страница 1: ...DIC C 268 DIC 162...

Страница 2: ...us testing manuals in Japan This model the Mupid 2plus retains all the performance of the Mupid 2 with improved usability based on the ideas of the researchers using it The Mupid 2plus can be used und...

Страница 3: ...icates the possibility of personal injury Follow these instructions completely Indicates the need to unplug the electrical cord Ignoring this instruction may cause an electric shock Indicates a testin...

Страница 4: ...tank lid is closed Attempting to do so can cause damage Do not put fingers or objects into the migration tank while electrical current is running Always make certain to turn off the power supply afte...

Страница 5: ...lid Power supply Migration tank 13P 4P 13 Migration tank Outside dimensions Migration tank internal dimensions Buffer capacity Material Weight Lid outside dimensions 187 5 mm W 139 5 mm L 57 0 mm H 1...

Страница 6: ...nt of sample is sufficient Make sure that the light source of the appropriate wave length for the stain is used If the gel is being observed in the migration tank or gel tray try observing the gel alo...

Страница 7: ...be apt to remain performing step 5 first is recommended 11P 6P 11 5 Product Maintenance Do not let buffer stand in the migration tank for a long time or re use it This can not only prevent accurate te...

Страница 8: ...lation may require more time if the agarose concentration is low 8 When the solution has sufficiently gelled hold the comb at each end and gently lift it upward Confirm the rectangular shape of the we...

Страница 9: ...espond to the pitch of a multipipette 1 2 5 After verifying that the operation switch of the power supply is OFF close the migration tank lid Do this as follows 1 First insert the 2 tabs on the cover...

Страница 10: ...espond to the pitch of a multipipette 1 2 5 After verifying that the operation switch of the power supply is OFF close the migration tank lid Do this as follows 1 First insert the 2 tabs on the cover...

Страница 11: ...lation may require more time if the agarose concentration is low 8 When the solution has sufficiently gelled hold the comb at each end and gently lift it upward Confirm the rectangular shape of the we...

Страница 12: ...be apt to remain performing step 5 first is recommended 11P 6P 11 5 Product Maintenance Do not let buffer stand in the migration tank for a long time or re use it This can not only prevent accurate te...

Страница 13: ...nt of sample is sufficient Make sure that the light source of the appropriate wave length for the stain is used If the gel is being observed in the migration tank or gel tray try observing the gel alo...

Страница 14: ...lid Power supply Migration tank 13P 4P 13 Migration tank Outside dimensions Migration tank internal dimensions Buffer capacity Material Weight Lid outside dimensions 187 5 mm W 139 5 mm L 57 0 mm H 1...

Страница 15: ...tank lid is closed Attempting to do so can cause damage Do not put fingers or objects into the migration tank while electrical current is running Always make certain to turn off the power supply afte...

Отзывы: