6

2 Gel 

Preparation

(1)  Place the gel maker stand level and set the gel tray into 

the stand. At maximum, four mini gels and two large 

gels can be prepared at once.

If the gel tray is not placed level, the thickness 

of the gel will not be uniform and migration 

may be affected.

If preparation is conducted at 80 

°

C or higher, 

there is a danger of deforming the gel maker 

stand and gel tray.

(2)  Weigh the required amounts of agarose powder and 

buffer solution in a conical flask or other vessels, and 

completely dissolve the agarose by applying heat.

(3)  Let the agarose solution stand until it cools to 70 

°

C or 

below.

(4)  Set the comb into the gel maker stand. Choose the 

small-toothed side (4 mm 

×

 8 or 17 wells) or the large-

toothed side (6 mm 

×

 6 or 12 wells) depending on the 

purpose. Multiple combs can also be set for a single gel.

(5)  Pour the appropriate amount of agarose solution into the 

gel tray. To prepare a gel of about 4 mm thickness, 

approximately 13 mL of agarose solution is required for 

a mini gel and 26 mL for a large gel.

The order of steps (4) and (5) can be 

reversed. For high-viscosity gels in which 

air bubbles may be apt to remain, 

performing step (5) first is recommended.

11P

6P

11

5    Product Maintenance

Do not let buffer stand in the migration tank for a long time or re-use it. This can 

not only prevent accurate test results, but can also cause damage.

Detach the power supply when discarding 

buffer or cleaning the migration tank.

When detaching the power supply, push the top 

with the thumb as illustrated in the figure at 

right. Don't pull on the electric cord.

If buffer is spilled on the power supply, dry it thoroughly and have it checked at a 

distributor as a precautionary measure.

Although the migration tank does not have any electrical switch, do not clean it by 

immersion in water.

Polishing or cleaning the electrode with a brush can cause breakage and should 

be avoided.

After cleaning, allow it to dry it in a place out of direct sunlight. Do not dry it under 

severe conditions such as a dry oven, or a hair dryer.

Do not clean this product with organic solvents such as acetone. Deformation 

could result.

Do not attempt to modify this product.

This product is designed to run on input of 100-120 V of alternating current; if 

using it in a country where voltage is higher, be sure to use a step-down 

transformer. When there is the necessity of keeping the temperature of buffer liquid 

low, It recommends using it in 100 VAC.

Содержание Mupid-2plus

Страница 1: ...DIC C 268 DIC 162...

Страница 2: ...us testing manuals in Japan This model the Mupid 2plus retains all the performance of the Mupid 2 with improved usability based on the ideas of the researchers using it The Mupid 2plus can be used und...

Страница 3: ...icates the possibility of personal injury Follow these instructions completely Indicates the need to unplug the electrical cord Ignoring this instruction may cause an electric shock Indicates a testin...

Страница 4: ...tank lid is closed Attempting to do so can cause damage Do not put fingers or objects into the migration tank while electrical current is running Always make certain to turn off the power supply afte...

Страница 5: ...lid Power supply Migration tank 13P 4P 13 Migration tank Outside dimensions Migration tank internal dimensions Buffer capacity Material Weight Lid outside dimensions 187 5 mm W 139 5 mm L 57 0 mm H 1...

Страница 6: ...nt of sample is sufficient Make sure that the light source of the appropriate wave length for the stain is used If the gel is being observed in the migration tank or gel tray try observing the gel alo...

Страница 7: ...be apt to remain performing step 5 first is recommended 11P 6P 11 5 Product Maintenance Do not let buffer stand in the migration tank for a long time or re use it This can not only prevent accurate te...

Страница 8: ...lation may require more time if the agarose concentration is low 8 When the solution has sufficiently gelled hold the comb at each end and gently lift it upward Confirm the rectangular shape of the we...

Страница 9: ...espond to the pitch of a multipipette 1 2 5 After verifying that the operation switch of the power supply is OFF close the migration tank lid Do this as follows 1 First insert the 2 tabs on the cover...

Страница 10: ...espond to the pitch of a multipipette 1 2 5 After verifying that the operation switch of the power supply is OFF close the migration tank lid Do this as follows 1 First insert the 2 tabs on the cover...

Страница 11: ...lation may require more time if the agarose concentration is low 8 When the solution has sufficiently gelled hold the comb at each end and gently lift it upward Confirm the rectangular shape of the we...

Страница 12: ...be apt to remain performing step 5 first is recommended 11P 6P 11 5 Product Maintenance Do not let buffer stand in the migration tank for a long time or re use it This can not only prevent accurate te...

Страница 13: ...nt of sample is sufficient Make sure that the light source of the appropriate wave length for the stain is used If the gel is being observed in the migration tank or gel tray try observing the gel alo...

Страница 14: ...lid Power supply Migration tank 13P 4P 13 Migration tank Outside dimensions Migration tank internal dimensions Buffer capacity Material Weight Lid outside dimensions 187 5 mm W 139 5 mm L 57 0 mm H 1...

Страница 15: ...tank lid is closed Attempting to do so can cause damage Do not put fingers or objects into the migration tank while electrical current is running Always make certain to turn off the power supply afte...

Отзывы: