7.7 Fuse Replacement
•Turn off power switch, remove power cord.
•Keep your hands dry.
①
Remove power cord from wall.
②
Open the cradle cover using a Philips
screwdriver as instructed on fuse cradle.
③
Replace with a new fuse (6GFU-F25A250V)
④
Reinstall the cradle cover.
⑤
Plug in the power cable.
⑥
Turn on the power to continue operation.
8 Daily Maintenance
To ensure a clean daily use, the mouthpiece hood should be cleaned frequently.
•To disinfect suction tubes, use VIROS OPTIM1 hydrogen peroxide disinfectant spray.
•To disinfect suction hood, avoid using disinfectant liquid that may change the shape or
color of the hood.
8.1 Maintenance Period of Equipment Parts
EOS Extraoral Suction System, Instructions For Use and Installation
20
Frequency
Content
Before Each Business Day
Check equipment surfaces and parts for any irregularities
On a Per Patient Basis
Suction hood Disinfection and Arms for interior Disinfection
After Each Business Day
Clean the equipment
Every 6 Months
Fine filter replacement
Every 12 Months
HEPA Filter Replacement
Every 12 Months
Noise filter stick replacement