7.3 Noise Filter Stick Replacement

replace time :12 months

•Cut off the main power and remove power cord.

•Keep your hands dry.

•Wear gloves.

[Caution] Replaced the noise filter stick and 

dispose of as medical waste.

 Loosen up the screws in the third joint and take 

out the noise filter stick.

 Place a new noise filter stick into the second 

suction arm.

 Insert the module of the third joint into the 

second suction tube. Tighten up the screws.

EOS Extraoral Suction System, Instructions For Use and Installation 

17

Fine Filter 

7.4 Fine Filter Replacement

replace time :6 months

•Turn off the main power and remove power cord 

from wall receptacle.

•Keep your hands dry.

•Wear gloves.

  [Caution] Replaced filters should be disposed of 

as medical waste.

 Loosen up the screws on door panel using a slot (flat 

head) type screwdriver. Remove the door panel.

 Unlock the dust-proof drawer cabinet, press the cover 

plate and remove the filter.

 Use a standard garbage bag to remove and dispose of 

the used fine filter as medical waste.

 Replace a new filter into the drawer.

 Replace the cover plate and reinstall the door panel.

 Replace particle/dust filter every 6 months.

Содержание EOS

Страница 1: ...http www adsequip com EOS Extraoral Suction System Instructions For Use and Installation P N 8027313 Version A Date of Issue 2020 05 06 Copyright ADS Dental System All Rights Reserved ...

Страница 2: ...terms and any regulations policies and guidelines established by ADS Understanding and agreeing to the terms of this disclaimer will hold ADS harmless to all personal injuries accidents property damage and legal disputes Except as stated in the after sales service policy all materials and contents related to the product are provided as things stand without any express or implied warranty and condi...

Страница 3: ...Instructions for replacing the suction hood 15 7 2 Instructions for replacing the suction arm 16 7 3 Instructions for replacing the noise filter stick 17 7 4 Instructions for replacing the fine filter 17 7 5 Instructions for replacing the HEPA filter 18 7 6 Instructions for replacing the UV Light 19 7 7 Instructions for replacing the fuse wire 20 8 Daily Maintenance 20 8 1 Maintenance Period of Eq...

Страница 4: ...d pathogens during dental procedures Do not use this product to take in water organic solvents Titanium powder or any solvent that are combustible This could lead to safety accidents Do not use this product to take in dirt sand rubbish etc Do not place this product anywhere near containers containing liquid especially hot liquid during use Keep the power lines away from sharp objects to avoid scra...

Страница 5: ...ibrillator In the case of any abnormalities in patients during use immediately turn off this product and discontinue use The electromagnetic wave from the product may cause cardiac pacemaker or cardioverter defibrillator malfunctions To avoid danger pay attention to the list below 1 The product should ONLY be operated or handled by dentists or by dental staff personnel under the supervision of a d...

Страница 6: ...the receptacle to avoid the dragging the power cord 6 Environment requirements The system should not be exposed to water The system should be free from possible hazards caused by pressure temperature humidity ventilation sunlight dust salt Sulphur containing air etc Avoid tilting or the bumping of the system when moving Never expose the system to chemicals or place the it near a chemical storage a...

Страница 7: ...r 14 8026746 Suction arm of the third joint 4 8026623 UV C Light 15 A121944 The third joint 5 8026563 HEPA filter 16 8026608 Noise filterstick 6 8027319 Transformer 17 A121943 Suction arm of the second joint 7 8026605 Castor 18 A121942 The second joint 8 8027343 Fuse 6GFU F25A110V 19 A121941 Suction arm of the first joint 9 8027340 Power cable 20 A121940 The first joint 10 8026686 Case 21 A121939 ...

Страница 8: ...filtration efficiency H14 matches European standard EN 1822 2009 ISO16890 and DOE STD 3020 2015 Specification for HEPA Filters Used by DOE Contractors Noise Decibel 58dB Tested under laboratory environment and 6 9 Inches distance from the suction mouth piece hood Suction Arm Caliber Φ2 UV Light Specifications Type UV C Lamp Tube Length 5 3 Lamp Tube Caliber 0 6 Lamp Cap Caliber 0 7 Wave Length 254...

Страница 9: ...4 1 Product Dimensions EOS Extraoral Suction System Instructions For Use and Installation 6 Unit inch ...

Страница 10: ...EOS Extraoral Suction System Instructions For Use and Installation 7 ...

Страница 11: ...tion arm Case Packing Size 27 2 x10 2 x9 14 6 x14 2 x39 4 Net Weight 3 2lbs 94 2lbs Gross Weight 5lbs 98 6lbs 5 Product Installation 1 After unpacking open the two door panels and remove the four fixed parts Warning If user does not take out the fixing parts to turn on the system which will cause motor damage ...

Страница 12: ...e screws provided Place tube through the hose clamp and connect it to the center post Then tighten the clamp 3 Insert the mounting hole of the first joint into the center post Make sure the thumb screw is aligned with the center pillar groove Then tighten the screw Caution Please install the suction arm from the end with a spring ...

Страница 13: ...e product to any source of light or heat either before during or after use To avoid possible bumping or tilting do not place objects on the cabinet base or near office furniture Never disassemble the joint cups before or during use This could lead to possible accidents or failure Do not tilt the equipment Otherwise it could lead to personnel injuries If the equipment is found in a tilted state do ...

Страница 14: ...Handle EOS Extraoral Suction System Instructions For Use and Installation Place the product on level ground Move by handles 11 ...

Страница 15: ...Ø Connect the 3 Prong Power Cord to the wall receptacle for 110V 20A power supply 110V 60Hz 20A EOS Extraoral Suction System Instructions For Use and Installation 6 2 Preparation 12 ...

Страница 16: ...ou can choose from To end or pause press the Start Stop button once Press it again to restart EOS Extraoral Suction System Instructions For Use and Installation 13 6 3 Turn On the Power Remove anything that might be inadvertently sucked in to the system Make sure the suction tubes are installed properly Press the Power button to turn the equipment on Suggestion After each dental treatment keep the...

Страница 17: ...he UVC lights are positioned on the HEPA filter and kills any remaining bacteria and viruses captured by the HEPA filter and are exhausted from the cabinet base When the suction stops the UVC light continually stays on for 30 minutes killing viruses pathogens and bacteria 6 5 Operation Principle Exhaust 6 6 Suction Arm Operation The rotating parts of the first joint center post is limited in its r...

Страница 18: ...nt 7 1 Suction mouth piece hood Daily Clean Use wiping disinfectant or spray disinfectant to clean the product surface ADS recommends to use VIROS OPTIM1 hydrogen peroxide disinfectant Replacement The suction mouthpiece hood is autoclavable Turn off the power and remove power cord Keep your hands Dry Wear gloves Loosen up the three screws in the suction mouth piece hood Replace with a new hood and...

Страница 19: ...ion tubes Dispose of as medical waste EOS Extraoral Suction System Instructions For Use and Installation 16 7 2 Suction Arms Daily Clean Use a spray disinfectant or wipes to clean the systems surface To internally clean the system turn it on to its lowest suction level spray VIROS OPTIM1 hydrogen peroxide disinfectant into the suction tubes and let the system run for 3 minutes Caution Do not disas...

Страница 20: ...tructions For Use and Installation 17 Fine Filter 7 4 Fine Filter Replacement replace time 6 months Turn off the main power and remove power cord from wall receptacle Keep your hands dry Wear gloves Caution Replaced filters should be disposed of as medical waste Loosen up the screws on door panel using a slot flat head type screwdriver Remove the door panel Unlock the dust proof drawer cabinet pre...

Страница 21: ...me Turn off the system power and remove powercord from the wall Keep your hands dry Wear gloves Remove the door panel Unlock the dust proof cabinet drawer and press the cover plate Remove the filter box Replace the HEPA filter Use a standard garbage bag to lift it from the device base and dispose of used HEPA filter as medical waste Replace the filter box replace the cover plate Double check and l...

Страница 22: ...ace of the HEPA filter Remove the UVC light tube and replace it with a new one dispose of the two as medical waste Remove the filter box and put the cover plate back on Double check and lock Reinstall the door panel Caution Replacement UV C Light should be disposed of as medical waste 19 UV C Light WARNING Do not look at UV lamps without using proper eye protection At all times wear gloves when ha...

Страница 23: ...on tubes use VIROS OPTIM1 hydrogen peroxide disinfectant spray To disinfect suction hood avoid using disinfectant liquid that may change the shape or color of the hood 8 1 Maintenance Period of Equipment Parts EOS Extraoral Suction System Instructions For Use and Installation 20 Frequency Content Before Each Business Day Check equipment surfaces and parts for any irregularities On a Per Patient Ba...

Страница 24: ... Should you need to service your system please contact the local dealer authorized by ADS n Email sales adsequip com Tel 626 6200456 n Technical Support 800 488 9708 Noise Filter Stick 8026722 HEPA Filter 8026563 Fuse 8027343 Fine Filter 8026725 Suction mouth piece hood 8026740 EOS Extraoral Suction System Instructions For Use and Installation 21 No Power Is the power on in the power socket Is the...

Страница 25: ...746 Suction arm of the third joint 4 8026623 UV C Light 15 A121944 The third joint 5 8026563 HEPA filter 16 8026608 Noise filterstick 6 8027319 Transformer 17 A121943 Suction arm of the second joint 7 8026605 Castor 18 A121942 The second joint 8 8027343 Fuse 6GFU F25A110V 19 A121941 Suction arm of the first joint 9 8027340 Power cable 20 A121940 The first joint 10 8026686 Case 21 A121939 Centre Po...

Страница 26: ...tion System use cables that are authorized by ADS only Attempt to use cable or other components from unauthorized source could lead to electromagnetic irregularities EOS Extraoral Suction System Instructions For Use and Installation 23 Statement of Manufacturer Electromagnetic Launch Extraoral Suction System is designed for electro magnetic environment described below Be sure to apply Lauch Test C...

Страница 27: ... wire to wire 2 kV ground to ground 1 kV wire to wire 2 kV ground to ground The power supply should reach the standard of typical commercial or hospital power supply Voltage sags short interruptions and voltage changes in power input lineGB T 17626 11 5 Ut lasting 0 5 Cycle At Ut 95 Sag 40 Ut lasting 5 Cycles At Ut 60 Sag 70 Ut lasting 25 Cycles At Ut 30 Sag 5 Ut lasting 5s At Ut 95 Sag 5 Ut lasti...

Страница 28: ...rmined by surveying the electromagnetic field a and in each frequency range dshould be lower than the compliance level Interference may occur near the equipment marked with the following symbol Note 1 At 80MHz and 800MHz frequencies apply higher frequency band formula Note 2 These guidelines may not be suitable for all situations Electromagnetic propagation is affected by absorption and reflection...

Страница 29: ...aximum output power Isolation distance corresponding to different frequencies of the transmitter m 150kHz 80MHz 80MHz 800MHz 80MHz 2 5GHz 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For the maximum rated output power of the transmitter not listed in the table above the recommended isolation distance d is in meters m which can be determined by the formula in the...

Страница 30: ...Keep Dry More Information Serial Number Grounding Connection This Way Up Caution Type B Machine Power Switch Start Stop Gear Up Gear Down UV Light 14 Symbols Descriptions EOS Extraoral Suction System Instructions For Use and Installation 27 ...

Страница 31: ...EOS Extraoral Suction System Instructions For Use and Installation 28 ...

Страница 32: ...00 488 9708 Tel 626 620 0456 E mail ads adsequip com www adsdental com www adsequip com Add 1590 S MILLIKEN AVE UNIT A ONTARIO CALIFORNIA 91761 USA P N 8027313 Version A Date of Issue 2020 05 06 Copyright ADS Dental System All Rights Reserved ...

Отзывы: