7.6
UV –C Light Replacement
EOS Extraoral Suction System, Instructions For Use and Installation
•Turn off the systems power and remove powercord
from wall.
•Keep your hands dry.
•Wear gloves.
①
Remove door panel.
②
Unlock the dust-proof cabinet drawer and replace
the cover plate.
③
Remove the filter box.
④
Turn over the filter box, making sure not to damage
the face of the HEPA filter.
⑤
Remove the UVC light tube and replace it with a
new one, dispose of the two as medical waste.
⑥
Remove the filter box and put the cover plate back
on. Double check and lock.
⑦
Reinstall the door panel.
[Caution]
Replacement UV –C
Light should be disposed of as
medical waste.
19
UV-C Light
[WARNING! ]
•
•Do not look at UV lamps without
using proper eye protection.
•
•At all times wear gloves when
handling UV lamps and quartz
sleeves, as oils from skin will
decrease UV energy transmission
once the lamp is heated and may
lead to premature failure.