Page 46
NodE 2
SET-POINT 2 OPERATING MODE
Select in sequence 4 operating criteria of set-point 2: [1410]
Comparison with net weight, with gross weight or with peak. In the latter case the comparison is carried
out with the last acquired peak value, even when the peak function is not active.
NET
The relay output is active in Net Weight mode. [0]
GROSS
The relay output is active in Gross Weight mode. [1]
PEAK
The relay output is active in Peak mode. [2]
PROCESS The relay output is active when the instrument is working normally. [3]
Default: GROSS
Selecting the output status if normally open or closed: [1411]
n. oPEn. Relay 2 is normally open. [0]
n.CLoSE Relay 2 is normally closed. [1]
Defaul: n. oPEn.
Select if positive or negative values have to be compared. [1412]
PoSIt..
The output is operative with positive weight. [0]
nEGAt.
The output is operative with negative weight. [1]
ALL:
The output is operative both with positive and negative weight. [2]
Default: PoSIt
Select whether only stable weight values are to be compared or also unstable: [1413]
norMAL Output 2 is active with unstable weight. [0]
StAbLE
Output 2 is active with stable weight. [1]
Default: norMAL
HySt. 2
SET-POINT 2 HYSTERESIS [1414]
Hysteresis value with respect to the set-point
Value: from 0 to Capacity
Default: 2
t1NEr2
SET-POINT 2 TIMING [1415]
Value of time, in tenths of a second, during which, when the weight value set is exceeded, the output
associated with set-point 2 remains enabled.
After this time, even if the weight value is still above the set-point, theoutput is automatically disabled.
The function is not active with programmed time equal to zero.
Values: from 000 to 999
Default: 0