V+ (10 kΩ min)
Ana-
mA+ (300 Ω max)
SHIELD
Page 9
The transmitter provides an analogue output in current or voltage.
Analogue output in voltage: range from 0 to 10 Volt or from 0 to 5
Analogue output in current: range from 0 to 20 mA or from 4 to 20
Analog transmission can be sensitive to electromagnetic
interference, it is therefore recommended that the cables
are as short as possible and that they follow their own route.
To make the connection, use a suitable shielded cable, ma-
king sure to connect the shield to the ground in one of the two ends.
Caution: do not connect the analogue output to active devices.
USB DEVICE
( SPECIFICATION 2.0 COMPLIANT; FULL-SPEED 12 MBPS)
Use this communication port to directly interface a PC via a USB port.
Use a standard USB cable for the connection.
To connect the instrument via the USB device, you must install a driver
on the PC which is suitable for the operating system used.
For installation please follow the specific instructions.
FIELDBUS CONNECTIONS
As an alternative to the RS485 serial port some of the most common
fieldbuses are available. You can use a single fieldbus which must
be specified when ordering.
ETHERNET CONNECTION
On the lower left part of the instrument there is a RJ45 connector
for Ethernet.
Features:
Trasmission speed 10 Mbps
Network compatible with 10/100/1000 Base-T
TCP Ethernet protocols, Modbus/TCP, UDP, IP, ICMP, ARP
TCP server communication mode
LED indicators (2) Presence of Ethernet and communication/dia-
gnostics line
Buffer size 256 bytes
Connection Time-out Min 30 seconds - Max 90 seconds
Link Time-out (cable disconnected) 30 seconds