2012-03-29 Manual Q-8V100 and S-25A30 Rev. 0.1
preliminary
25
Code
Description
0
All tests passed; no error found
1
FPGA boot failed
2
FPGA not running
4
Configuration data corrupt
8
Setting data corrupt
128
Output look-up table data corrupt
256
Sensor data alignment error
6.8.2. Error diagnosis
Since the communication protocol does not check the message content, the error result of a
command can be obtained by a separate command.
Command Syntax:
ERR?
In case of no error, a 0 is returned. The returned values for various errors are listed below.
Code
Description
1
Last received command: not recognized.
2
Last received command: missing parameter(s)
3
Last received command: parameter syntax error
4
Last received command: too many parameters
5
Last received command: not enough parameters
7
One or more of the command parameters for the last received command is
out of range.
8
Last received command: internal Adimec error
100
Loading of settings from non-volatile memory failed
101
Storing of settings to non-volatile memory failed
102
The defect pixel correction data can’t be stored, because the defect pixel list
is full
103
The defect pixel correction data can’t be stored, because the defect pixel is
already in the defect pixel list
120
LUT begin action not allowed; LUT transactions already pending
121
Adding LUT elements not allowed
122
Adding LUT elements stopped before complete LUT was filled
123
Adding more LUT entries than allowed
124
The current state does not allow the action
6.8.3. Cross hairs and test pattern
Cross hairs and test pattern are available in the camera for aligning the image and
debugging the imaging pipeline.
Command Syntax:
TPx
TP?
Where x is as follows: