84
2. Task Execution
:
Press to execute the task
:
1). Press to zero pressure module
if necessary
2). Press to start the test;
◆
The calibrator start pressurizing to the first test point
◆
Press
to switch calibration interface
3). Standard test steps:
(1). After pressure is stable , enter the UUT pressure value on the top right UUT window or press
to switch the calibration interface.
(2). Press the Enter key or
for the next test point in which the calibrator will pressurize again.
Note: if the tapping function is enabled, you should tap the dial gauge at each set point.
(3). After the next pressure is stable then enter the UUT value in the UUT window or press
to switch
the calibration interface.
(4). Repeat step (2) and (3) until all test points are completed.
3. Report:
Once test is done,
the reporting interface will appear. For more information please see section 6.4.
www.
.com
1.800.561.8187
Содержание ADT760
Страница 1: ...www com information itm com 1 800 561 8187...
Страница 2: ...I www com information itm com 1 800 561 8187...
Страница 12: ...XI Figure 19 REF and Outlet Port Filter 98 www com information itm com 1 800 561 8187...
Страница 18: ...6 1 2 Basic Structure Figure 1 Basic Structure www com information itm com 1 800 561 8187...
Страница 28: ...16 Figure 2 Standard Equipment www com information itm com 1 800 561 8187...
Страница 35: ...23 2 5 Pressure Connection Figure 6 Pressure Connection www com information itm com 1 800 561 8187...
Страница 40: ...28 2 Connection Figure 8 Output Connection www com information itm com 1 800 561 8187...
Страница 148: ...136 Step 4 Click on Manual Install www com information itm com 1 800 561 8187...
Страница 158: ...146 Step 2 Select No to skip update process www com information itm com 1 800 561 8187...