3.3 Zeroing
The zeroing function is available for current/voltage measurement and pressure measurement either
internal or external pressure modules.
◆
To zero: Press
and select
◆
Apply a short circuit for current / voltage zeroing
◆
Allowable zeroing range:
Pressure: within 1%FS;
Current / Voltage: within 0.1%FS.
3.4 Vent
Press Vent on Control Bar and the calibrator will run the vent process.
◆
The vent process will end pressurization during a regular test, typical application, task, and leak
test
◆
The VENT port should not face the operator during venting process
◆
Dust or contaminants may blow out through the VENT port
◆
Any moisture or liquid may also blow out through the vent port