53
See section 3.1.3-3-
③
Auto Step.
2.2 Test Execution:
After test settings are complete, press to execute the test:
1). Press to zero pressure module if necessary;
2). Press to start the test;
◆
The calibrator will pressurize to the first set point (which may be zero)
◆
Press
to switch calibration interface
◆
Press
to check or change test settings
3). Standard test steps:
(1). After the pressure shows it is stable, the gauge pressure value can be entered in the box on
the right. Press
to switch the calibration interface.
(2). Press the enter key or
for next test point and the calibrator will start to pressurize again
(3). Wait for the pressure value to be generated and stabilize and then enter the UUT values and
proceed to the next point.
(4). Repeat step (2) and (3) until all test points passed.
2.3 Report:
Once the test is complete, the test report interface will appear. For more information, please see
section 4.6.
www.
.com
1.800.561.8187
Содержание ADT760
Страница 1: ...www com information itm com 1 800 561 8187...
Страница 2: ...I www com information itm com 1 800 561 8187...
Страница 12: ...XI Figure 19 REF and Outlet Port Filter 98 www com information itm com 1 800 561 8187...
Страница 18: ...6 1 2 Basic Structure Figure 1 Basic Structure www com information itm com 1 800 561 8187...
Страница 28: ...16 Figure 2 Standard Equipment www com information itm com 1 800 561 8187...
Страница 35: ...23 2 5 Pressure Connection Figure 6 Pressure Connection www com information itm com 1 800 561 8187...
Страница 40: ...28 2 Connection Figure 8 Output Connection www com information itm com 1 800 561 8187...
Страница 148: ...136 Step 4 Click on Manual Install www com information itm com 1 800 561 8187...
Страница 158: ...146 Step 2 Select No to skip update process www com information itm com 1 800 561 8187...