45
3. HART Device Output
Subject
Type
Valid Value
Comment
PV / Range Unit
Editable
Select from 25 default units
or from 5 custom units
HART device pressure measurement
unit, see section 3.1.2
PV LRV
Editable
Within HART device sensor
range
Process variable lower range
PV URV
Editable
Within HART device sensor
range
Process variable upper range
Transfer function
Editable
Linear / Square Root
Transfer function type
Alarm State
Read Only
N/A
Alarm State
Damping
Editable
1~100
Damping time, only support one digit
after decimal point
Polling Address
Editable
0~15
HART device polling address
Burst Mode
Editable
On / Off
Burst mode status
Burst Command
Editable
1 / 2 / 3
Burst command, depends on HART
device
Table 16 HART Device Output
www.
.com
1.800.561.8187
Содержание ADT760
Страница 1: ...www com information itm com 1 800 561 8187...
Страница 2: ...I www com information itm com 1 800 561 8187...
Страница 12: ...XI Figure 19 REF and Outlet Port Filter 98 www com information itm com 1 800 561 8187...
Страница 18: ...6 1 2 Basic Structure Figure 1 Basic Structure www com information itm com 1 800 561 8187...
Страница 28: ...16 Figure 2 Standard Equipment www com information itm com 1 800 561 8187...
Страница 35: ...23 2 5 Pressure Connection Figure 6 Pressure Connection www com information itm com 1 800 561 8187...
Страница 40: ...28 2 Connection Figure 8 Output Connection www com information itm com 1 800 561 8187...
Страница 148: ...136 Step 4 Click on Manual Install www com information itm com 1 800 561 8187...
Страница 158: ...146 Step 2 Select No to skip update process www com information itm com 1 800 561 8187...