5

IF A MERCURY SPILL OCCURS . . .

FAQ (FREQUENTLY ASKED QUESTIONS)

Why use mercury?

Mercury (Hg) is a liquid metal with very stable performance charac-
teristics. As such, mercurial blood pressure devices remain the interna-
tionally recognized standard for accuracy. Because there are no moving
parts, springs, needles, casings, or other components that may become
defective or wear with age,  mercury instruments are inherently the
most reliable and accurate. With minimal care and maintenance, these
instruments should last a lifetime. 

What is so special about DIAGNOSTIX™ mercury instruments?

Our exclusive Safety First

system incorporates a number of distinctive

safety features that provide you with a level of protection no other
mercurial instrument can match.

Unbreakable graduated plastic cartridge tube

. Designed from space age

polymers, the tube exhibits remarkable performance stability. It will
not crack, craze, or shatter. Yellowing, a problem that plagued first
generation plastic tubes, is all but gone. Tolerances to .1mm assure
measurement repeatability. 

Reservoir Locking Switch

secures the mercury safely in the reservoir

during routine maintenance, cleaning, or transport.

EZ Tube

System

simplifies the safe removal of the cartridge tube or

top cap filters for cleaning or maintenance without risking a spill. 

4. To replace tube, lift EZ-Tube

mechanism and insert base of

tube into main unit, being sure graduation marks face out. Next,
position top of tube and release EZ-Tube

mechanism. On 922 re-

place by sliding down through top cap opening and position. Se-
cure top cap by turning clockwise until tight. Be sure to tighten
thoroughly.

5. Unlock reservoir and tilt instrument AWAY from reservoir forcing
mercury back into tube. 

6. Check for “0” point.

Содержание Diagnostix 720

Страница 1: ...JUSTINCASE ADC GUIDELINES for the Proper Maintenance Handling Cleanup and Transportation of Mercurial Blood Pressure Devices...

Страница 2: ...future reference PREVENTIVE MAINTENANCE Because the mercury units do not rely on moving parts such as metal bellows diaphragms gears or other sources of mechanical wear only an occasional inspection o...

Страница 3: ...n tipped in the direction of the reservoir and returned to a vertical position before the new mercury level is observed Re peat the process as necessary to bring the mercury up to the 0 cali bration m...

Страница 4: ...e mer cury has been lost add enough through the top of the plastic tube to bring the rim of the mercury meniscus to the 0 calibration mark Then replace the knurled cap at the top of the graduated plas...

Страница 5: ...ed plastic cartridge tube Designed from space age polymers the tube exhibits remarkable performance stability It will not crack craze or shatter Yellowing a problem that plagued first generation plast...

Страница 6: ...lled mercury should be picked up at once since the mercury will vaporize rapidly and may contaminate the surrounding air Larger spills involving the entire contents of the mercury instrument about 2 o...

Страница 7: ...do not have one they can be ordered from most laboratory safety supply houses As a conven ience to our customers a basic spill kit can be purchased directly from ADC Call our customer service departme...

Страница 8: ...What do I do with the collected mercury and broken instrument You have two options You can dispose of the collected mercury and damaged instrument in accordance with your state and local environ menta...

Страница 9: ...n below Note We can only accept the return of ADC s DIAGNOSTIX mer curial instruments for repair services PACKING AND SHIPPING INSTRUCTIONS Important Note Failure to follow the proper packing and ship...

Страница 10: ...osed with a strip of adhesive tape Carefully place the specially authorized return address label to the top of the box 9 Ship via UPS Prepaid Ground Service ONLY No other carriers are authorized use f...

Страница 11: ...ange top cap filter Clean or replace mercury Add mercury Clean tube in accordance with en closed instructions Change top cap filter Make sure cartridge tube is securely seated Replace silicone gaskets...

Страница 12: ...follow the links Consult this manual for basic service and maintenance procedures mentioned in this troubleshooting guide If any of the problems persist contact ADC s Customer Service Department Toll...

Отзывы: