background image

11

Air Bubble in car-

tridge tube during

inflation                 

                             

                             

Blood Pressure

Reading seems

High                      

                             

                             

                             

Blood Pressure

Reading seems

Low

Dirty cartridge

tube                      

                             

Mercury bouncing

in tube during in-

flation or deflation

Mercury drops

appear outside

the top or bottom

of the cartridge

tube

Mercury rises or

falls slowly            

                             

                             

Mercury sepa-

rates in tube dur-

ing deflation          

• Add mercury by following the in-

structions in this manual 

• To remove trapped air, tap reservoir

while system is inflated

• Be sure to deflate the inflation sys-

tem at the recommended rate of 2-

3mmHg/sec

• Change top cap filter

• Clean or replace mercury

• Add mercury

• Clean tube in accordance with en-

closed instructions

• Change top cap filter

• Make sure cartridge tube is securely

seated

• Replace silicone gaskets at top and

bottom of tube

• Change top cap filter

• Clean tube in accordance with en-

closed  instructions

• Be sure to deflate the inflation sys-

tem at the recommended rate of 2-

3mmHg/sec 

• To remove trapped air, tap reservoir-

while system is inflated

• Insufficient

amount of mercury

• Air trapped in

reservoir

• Excessive deflation

rate

• Air flow blockage

• Mercury level low

• Mercury has oxi-

dized leaving a silver

gray residue

• Filters need re-

placements

• Leak at seals

• Air flow blockage

• Excessive deflation

rate

• Trapped air

                                

      PROBLEM              LIKELY  CAUSE                           RESOLUTION

Содержание Diagnostix 720

Страница 1: ...JUSTINCASE ADC GUIDELINES for the Proper Maintenance Handling Cleanup and Transportation of Mercurial Blood Pressure Devices...

Страница 2: ...future reference PREVENTIVE MAINTENANCE Because the mercury units do not rely on moving parts such as metal bellows diaphragms gears or other sources of mechanical wear only an occasional inspection o...

Страница 3: ...n tipped in the direction of the reservoir and returned to a vertical position before the new mercury level is observed Re peat the process as necessary to bring the mercury up to the 0 cali bration m...

Страница 4: ...e mer cury has been lost add enough through the top of the plastic tube to bring the rim of the mercury meniscus to the 0 calibration mark Then replace the knurled cap at the top of the graduated plas...

Страница 5: ...ed plastic cartridge tube Designed from space age polymers the tube exhibits remarkable performance stability It will not crack craze or shatter Yellowing a problem that plagued first generation plast...

Страница 6: ...lled mercury should be picked up at once since the mercury will vaporize rapidly and may contaminate the surrounding air Larger spills involving the entire contents of the mercury instrument about 2 o...

Страница 7: ...do not have one they can be ordered from most laboratory safety supply houses As a conven ience to our customers a basic spill kit can be purchased directly from ADC Call our customer service departme...

Страница 8: ...What do I do with the collected mercury and broken instrument You have two options You can dispose of the collected mercury and damaged instrument in accordance with your state and local environ menta...

Страница 9: ...n below Note We can only accept the return of ADC s DIAGNOSTIX mer curial instruments for repair services PACKING AND SHIPPING INSTRUCTIONS Important Note Failure to follow the proper packing and ship...

Страница 10: ...osed with a strip of adhesive tape Carefully place the specially authorized return address label to the top of the box 9 Ship via UPS Prepaid Ground Service ONLY No other carriers are authorized use f...

Страница 11: ...ange top cap filter Clean or replace mercury Add mercury Clean tube in accordance with en closed instructions Change top cap filter Make sure cartridge tube is securely seated Replace silicone gaskets...

Страница 12: ...follow the links Consult this manual for basic service and maintenance procedures mentioned in this troubleshooting guide If any of the problems persist contact ADC s Customer Service Department Toll...

Отзывы: