10

2. Put ONE mercurial unit in a polybag and seal it with a twist tie
or rubber band closure.

3. Wrap bubble wrap around the mercurial unit and secure with
adhesive tape. This will help prevent the port connector from
puncturing the second polybag.

4. Enclose the wrapped mercurial unit in another polybag. Seal
the polybag with another twist tie or rubber band closure.

5. Place the double polybagged and wrapped unit inside the
packing case,or inner carton.

6. Arrange some cushioning bags or more bubble wrap around
the mercurial unit to protect it from excessive shifting.

7. If returning two or three mercurial units, repeat steps 1-6 as
necessary. We suggest packaging each of the mercurial units in
separate inner cartons that can fit into the packing case.

8. Seal the packing case closed with a strip of adhesive tape.
Carefully place the 

specially authorized

return address label to

the top of the box.

9. Ship via UPS Prepaid Ground Service ONLY. No other carriers
are authorized use for safe return of mercury instruments.

What is a Safety Data Sheet (SDS)? Why do I need it and how
do I get one?

An SDS sheet is required by OSHA. All medical locations require you
keep it on file. An SDS provides you with information about mercury,
and what protective measures you have to do for your safety. Your
ADC dealer will provide you with the SDS upon request. 

The information contained within this guide is provided in good faith
and is believed to be correct as of the date published. ADC makes no
representation as to the completeness or accuracy of the information.
It  is  understood  that  the  person  receiving  this  information  will  use
good judgement to determine proper actions to be taken. Information
contained within does not provide any warranties, expressed or im-
plied.

Содержание Diagnostix 720

Страница 1: ...JUSTINCASE ADC GUIDELINES for the Proper Maintenance Handling Cleanup and Transportation of Mercurial Blood Pressure Devices...

Страница 2: ...future reference PREVENTIVE MAINTENANCE Because the mercury units do not rely on moving parts such as metal bellows diaphragms gears or other sources of mechanical wear only an occasional inspection o...

Страница 3: ...n tipped in the direction of the reservoir and returned to a vertical position before the new mercury level is observed Re peat the process as necessary to bring the mercury up to the 0 cali bration m...

Страница 4: ...e mer cury has been lost add enough through the top of the plastic tube to bring the rim of the mercury meniscus to the 0 calibration mark Then replace the knurled cap at the top of the graduated plas...

Страница 5: ...ed plastic cartridge tube Designed from space age polymers the tube exhibits remarkable performance stability It will not crack craze or shatter Yellowing a problem that plagued first generation plast...

Страница 6: ...lled mercury should be picked up at once since the mercury will vaporize rapidly and may contaminate the surrounding air Larger spills involving the entire contents of the mercury instrument about 2 o...

Страница 7: ...do not have one they can be ordered from most laboratory safety supply houses As a conven ience to our customers a basic spill kit can be purchased directly from ADC Call our customer service departme...

Страница 8: ...What do I do with the collected mercury and broken instrument You have two options You can dispose of the collected mercury and damaged instrument in accordance with your state and local environ menta...

Страница 9: ...n below Note We can only accept the return of ADC s DIAGNOSTIX mer curial instruments for repair services PACKING AND SHIPPING INSTRUCTIONS Important Note Failure to follow the proper packing and ship...

Страница 10: ...osed with a strip of adhesive tape Carefully place the specially authorized return address label to the top of the box 9 Ship via UPS Prepaid Ground Service ONLY No other carriers are authorized use f...

Страница 11: ...ange top cap filter Clean or replace mercury Add mercury Clean tube in accordance with en closed instructions Change top cap filter Make sure cartridge tube is securely seated Replace silicone gaskets...

Страница 12: ...follow the links Consult this manual for basic service and maintenance procedures mentioned in this troubleshooting guide If any of the problems persist contact ADC s Customer Service Department Toll...

Отзывы: