CLEARPASS
Full-Body Security Screening System
OPERATION
MANUAL
Identification
ADN190.00.00.000 OM2
Revision
1
Date
02.2022
Page
23 of 52
2.4 ELECTRICAL SAFETY
The System complies with the electrical safety requirements established in Low Voltage
Directive 2014/35/EU and technical standard EN 61010-1:2010.
By the type of protection against electric shock the System refers to Class I accord-ing
to EN 61010-1:2010.
The System design in the part of the conductive part insulation, gap width and leakage
paths meet the requirements of the installation category (overvoltage category) II with a pollu-
tion degree 1 according to EN 61010-1:2010.
The design of the System ensures protection against contact with the energized parts
and with those parts which can become energized in case of a single fault.
2.4.1 Electromagnetic compatibility
The System meets the requirements for electromagnetic compatibility requirements es-
tablished in IEC 61326-1:2013, EN 55011:2016 and Electromagnetic compatibility Directive
2014/30/
Е
U.
The System belongs to the equipment of class A according to EN 55011:2016.
The protective covers of the Scanner have IP20 protection level in accordance with IEC
60529.
IT IS FORBIDDEN TO SUPPLY ELECTRIC POWER TO THE SYS-
TEM UNTIL RELEVANT INSTRUCTIONS IN THIS MANUAL ARE
ACQUIRED.
DO NOT REMOVE THE PROTECTIVE COVERS OF THE SYSTEM
OR PERFORM ANY OPERATIONS ON THE INTERNAL PARTS
WHEN THE SYSTEM IS ON. IT CAN CAUSE SERIOUS HEALTH
DAMAGE AND SYSTEM DAMAGE.
The System is not intended for use in residential environments
and may not provide adequate protection to radio reception in
such environments.
Difficulties may occur in ensuring electromagnetic compatibility
in other areas due to conductive and emitted interference.