9.0 RS-232 INTERFACE
The AGF floor scale is supplied with bi-directional RS-232 interface as standard. The scale
when connected to a printer or computer outputs the weight with the selected weighing
unit through the RS-232 interface.
Specifications:
RS-232 serial interface is a plug as figure 2 shows:
1: Pin GND, Signal Ground
2: Pin RXD, Received Data
3: Pin TXD, Transmitted Data
As viewed from the back of the indicator
The scale can be set to print text in English, French, German, Spanish, Italian or
Portuguese. See the RS-232 parameters section for details.
DATA FORMAT - COMPLETE WEIGHT
<cr><lf>
<cr><lf>
Date
12/09/2006 <cr><lf>
Time
14:56:27 <cr><lf>
<cr><lf>
Scale ID
123456 <cr><lf>
If ID is zero, it is left blank
User ID
234567 <cr><lf>
<cr><lf>
Net Wt
1.234 Kg <cr><lf>
Net Wt. (or Gross Wt.)
Tare Wt.
0.000kg
Gross Wt.
1.234 Kg
<cr><lf>
<cr><lf>
<cr><lf>
<cr><lf>
RS‐232 output of weighing data
ASCII code
9600 Baud (user selectable)
8 data bits
No Parity