EN
© Adam Equipment Company 2015
35
3016612481Rev2.00-Jul15
larger number corresponds to a larger stable zone. Selectable values: 1, 2, 5
or 10 (divisions).
Recommended value for normal use: 1
AV ZERO
AV ZERO
AV ZERO
AV ZERO
ON = Auto-zero function. Selectable values: 1, 2, 5, 10 or 15 (divisions).
OFF = Auto-zero function disabled.
Recommended value for normal use: ON, 5
SEPARAT
SEPARAT
SEPARAT
SEPARAT
COMMA Set separator indicator on the display to be either a decimal point
DEC PT or a comma. Also applies to the serial interface for print output.
13.6 CALIBRATION SETUP
This menu allows the Supervisor to set the calibration parameters. Press [Setup] when
“
CAL SET
CAL SET
CAL SET
CAL SET
” is displayed to select the calibration parameters. The options for each
parameter can be scrolled through by using the [Up] or [Down] key and pressing [Setup]
to confirm choices.
ENABLE
ENABLE
ENABLE
ENABLE
NO = Operator calibration is disabled.
YES = Operator calibration is enabled.
CAL REP
CAL REP
CAL REP
CAL REP
ON = Prints out Calibration report after successful calibration.
OFF = Disabled.
TIM CAL
TIM CAL
TIM CAL
TIM CAL
ON = Enabled. Select time from 1 to 24 hours.
OFF = Disabled.
TEM CAL
TEM CAL
TEM CAL
TEM CAL
ON = Enabled. Select the temperature variation from 0.2 to 4°C which when
detected will trigger automatic calibration.
OFF = Disabled.
INT CAL
INT CAL
INT CAL
INT CAL
YES = Internal calibration enabled (if fitted).
NO = External calibration enabled.
INT MAS
INT MAS
INT MAS
INT MAS
CAL MAS = Displays the set value of the internal calibration mass (if fitted) in
grams. If after verification against an external mass it is determined that the
value of the internal mass needs adjustment, e.g. due to wear, accumulation
of dirt, etc., then this value can be adjusted by +/- 100 mg. This should only
be considered by expert users if the external reference weight is definitively
accurate and an incorrect weight reading is being given after internal
calibration. Adjustment will restore the internal calibration to the correct level
of accuracy.
Press [Mode] to advance to setting of the next menu or [
0/T
] to return to normal
weighing.
13.7 PASSCODES
To enable the security features in this balance it is necessary to set passcodes. There are
2 passcodes called Operator Passcode and Supervisor Passcode. The Operator
Passcode allows an authorised user to operate the basic weighing functions of the balance
but will not allow access to the Supervisor Menus if the Supervisor Passcode has been
set.
Note: To change or disable a Passcode it is be necessary to enter the current passcode.
To setup passcodes:
Press [Setup]. Use the [Up] and [Down] keys to cycle through options until “
PASSCOd
PASSCOd
PASSCOd
PASSCOd
” is
displayed. Press [Setup] again to enter this section. Use [Up] and [Down] keys to select
Operator (“
PC OPER
PC OPER
PC OPER
PC OPER
”) or Supervisor (“
PC SV
PC SV
PC SV
PC SVPE
PE
PE
PE
”) option.