PROCEDURE
•
While in the normal weighing condition, press and hold the
[Tare/Zero]
key for 4 seconds.
•
The display will show a “
CAL
” along with the last selected unit. The
unit can be changed by using the
[Unit]
key to calibrate in kg or lb.
•
Press
the
[Print/Hold]
key. The display will show “
L xx
” where
xx
is
the Calibration weight which is user-selectable.
•
Use the
[Tare/Zero]
key to change the flashing digit and the
[Print/Hold]
key to move to the next digit.
•
Press
the
[Unit]
key to confirm the calibration weight. The digit stops
flashing.
•
If the selected mass is less than 10% of the capacity of the scale, an
error message “
CALEr
” will be displayed and the scale will return to
zero. Repeat the process correctly.
•
Place the correct calibration mass as selected by the user at the
centre of the pan.
•
Press
the
[Unit]
. The display will return to weighing mode.
•
If the mass loaded is more than ±20% of the factory calibration
reference then an error message “
CALEr
” will be displayed and the
scale will return to weighing without calibration being saved. Repeat
the process correctly.
•
Remove the weight.
•
Verify the scale is calibrated correctly. If not then check the scale
and repeat the process.