© Adam Equipment Company 2006
22
11.0 ERROR CODES
During the initial power-on testing or during operation, the scale may show an
error message. The meaning of the error messages is described below.
If an error message is shown, repeat the step that caused the message. If the
error message is still shown then contact your dealer for support.
ERROR CODE
DESCRIPTION
POSSIBLE CAUSES
Err 4
Initial Zero is greater than
allowed (4% of maximum
capacity) when power is turned
on or when the
[Zero/Enter]
key
is pressed.
Weight on the pan when turning
the scale on.
Excessive weight on the pan when
zeroing the scale.
Improper calibration of the scale.
Damaged load cell.
Damaged Electronics.
Err 5
Keyboard error.
Improper operation of the scale.
Err 6
A/D count is not correct when
turning the scale on.
Platform is not installed.
Load cell is damaged.
Electronics is damaged.
Err 7
Scale is not stable when setting
the percentage weighing.
The scale is not stable.
Improper operation of the scale.
FAIL H
or
FAIL L
Calibration
error.
Improper calibration (should be
10% of the factory
calibration). The old calibration
data will be retained until the
calibration process is complete.
Err 9
Scale is unstable.
There is vibration or draft making
the scale unstable.
Electronics may be damaged.
To view the A/D count, press
[Zero/Enter]
and
[%]
at the same time while the
scale is performing the initial check at power-on. Press
[Zero/Enter]
to return
to normal weighing.
Содержание CBW 15
Страница 2: ...Adam Equipment Company 2006...
Страница 4: ...Adam Equipment Company 2006 2...