10.0 CALIBRATION
•
The
CBW
scales are calibrated using metric weights.
•
To start calibration turn the scale off and then turn it on again.
Press the
[Tare]
and
[%]
keys together once, during the initial
counting from 9 to 0 on the display.
•
The display will show "
UnLoAd
" for all models except CBW 6H.
Remove all weight from the pan and then press the
[Zero/Enter]
key when the scale is stable.
Note: For CBW 6H,
it will show
"
LoAd 0
" and when the scale is stable, it will ask for the calibration
mass to be placed on the platform. For all other models, the
display will show “
AdLoAd
”. Place the suggested calibration mass
on the pan. It is best to use a weight close to the full capacity of
the scale.
•
Press the
[Zero/Enter]
key. The display will then show the last
mass used. If this is same as the mass on the pan then it can be
used or a different value can be entered.
•
To use this value, press the
[Zero/Enter]
key when the stable
indicator is on.
•
If a different value is desired, enter the value using the
[
►
]
arrow
key to select the flashing digit and the
[
▲
]
to increment the
flashing digit. If value is correct, press the
[Zero/Enter]
key to
accept when the stable indicator is on.
•
Remove the calibration weight as the scale counts back from 9 to
0.
•
If an error message “
FAIL H
” or “
FAIL L
” is shown during calibration,
re-check the calibration and repeat, if necessary. If the error
cannot be corrected contact your dealer or Adam Equipment for
advice.