This  will cause the bridge to become unbalanced.   When the bridge  has a voltage 

across it from +E to -E then the signal output at +S and -S will show a voltage of zero 

volts  with  no  load  and  a  small  voltage  proportional  to  the  load  as  the  load  is 

increased.      Typical  load  cells  show  20-30mv  of  signal  if  the  excitation  voltage  is 

10volts and the load cell is fully loaded. 

 

Most  load  cells  used  in  scales  are  not  as  simple  as  this  example.    They  have 

additional resistance elements added to compensate for temperature variations and 

to  set  the  outputs  to  correct  voltages.    The  circuit  of  a  typical  load  cell  is  shown 

below-  

 

 

 

A method to check the basic function and integrity of a load cell can be done using a 

ohmmeter and voltmeter with up to a 10volt power supply. 

 

Before the load cell is connected to the power supply use the ohmmeter to measure 

the resistance between the wires.  Typical resistance values are: 

 

+E to -E 

 

410 ohms  ±30 ohms 

 

+S to -S 

 

350 ohms  ±  2 ohms 

 

+E to +S or any other combination similar approximately 270 ohms. 

 

If  the  load  cell  has  Sense  connections  in  addition  they  are  connected  to  the 

excitation internal to the load cell. 

 

Connect  the  load  cell  +E  and  -E  to  a  power  supply,  typically  10  volts.    Never  use 

more than 12 volts as it may damage the load cell. 

Measure the voltage bS and -S.   

 

With no load the voltage should be approximate 0mv ±3mv. 

 

With  a  mass  on  the  load  cell  the  output  voltage  should  increase.    The  amount  by 

which it increases will be dependent upon the sensitivity of the load cell, capacity of 

the load cell, the excitation voltage and the amount of mass placed on the load cell.  

Most load cells have a sensitivity of either 2mv/V or 3mv/V. 

Содержание CBW 100a

Страница 1: ...Adam Equipment Company 2005 Adam Equipment CBW SERIES SERVICE MANUAL ADAM EQUIPMENT CO LTD p n 4331 Rev A3 January 2005...

Страница 2: ...0 ERROR CODES 5 4 0 CBW DESCRIPTION 6 5 0 CBW CALIBRATION PARAMETERS VER 1 2 2 03 7 5 1 CBW CALIBRATION VER 1 2 2 03 7 5 2 CBW PARAMETERS VER 1 2 2 03 8 6 0 TWO POINT FACTORY CALIBRATION 10 6 1 ADC OU...

Страница 3: ...is blank Power Switch faulty or not connected No turn on test Battery not charged or faulty Power supply faulty or incorrect type E6 is displayed Pan not installed Poor connection Load cell damaged El...

Страница 4: ...te weight Balance not level Obstruction between sample pan Wrong unit of weight displayed Cannot use Full Capacity Over load Stops hitting pan support or hitting bottom of load cell Electronics faulty...

Страница 5: ...Keyboard error Improper operation of the scale A D count is not correct when turning the scale on Platform is not installed Load cell is damaged Electronics is damaged Scale is not stable when settin...

Страница 6: ...W are similar except the selection of load cells and the program The main circuit board is the same as used in the CBC scales Normally if a problem is found with a circuit board the most cost effectiv...

Страница 7: ...ero Enter key when the stable indicator is on Or if a different value is desired enter the value using the arrow key to select the flashing digit and the to increment the flashing digit If value is co...

Страница 8: ...Print keys at the same time Setting the scale increment The display will show where can be 1 2 5 10 or 20 The first value shown is the default scale increment value To change to a different increment...

Страница 9: ...division The display will show Press the Tare key to change the value Press the Zero Enter key to store the changed setting Selectable unit of weight displayed The scales are normally set to display...

Страница 10: ...cator will not show or when the raw counts do not seem to be moving Please ensure that the load cell is not damaged and that the load cell is wired correctly with the right output voltages To start th...

Страница 11: ...ead load percentage of the cell If with no load on the pan the counts are below 1000 or above 200 000 then the load cell should be tested as explained in the next section Apply full capacity to the pa...

Страница 12: ...articular way on the metal structure These 4 strain gauges are connected in such a way that they form a wheatstone bridge See figure below In the simplest load cells there are only the 4 resistances o...

Страница 13: ...voltmeter with up to a 10volt power supply Before the load cell is connected to the power supply use the ohmmeter to measure the resistance between the wires Typical resistance values are E to E 410...

Страница 14: ...needs to be amplified to make it acceptable for the A D converter used The TCC CBW scales amplify the input by 200 and change the no load voltage to about 2 5V The TFC and TFW scales use a gain of 500...

Страница 15: ...he top pan and take out the 5 screws from the bottom of the scale 2 Detach the 4 cables connected with the Load cell Display Power Switch and Battery Chamber from the main PCB Remove the main PCB by t...

Страница 16: ...ale on to verify proper installation Set the parameters as described in the Service Manual and perform the Factory Calibration by using the Limit and keys together If the RS 232 needs to be disabled o...

Страница 17: ...Adam Equipment Company 2005 17...

Страница 18: ...Adam Equipment Company 2005 18...

Страница 19: ...to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment The equipment generates uses and can radiate radio frequency energy and if not...

Страница 20: ...tronic Scales for Medical use Retail Scales for price computing Adam Equipment Co Ltd Bond Avenue Milton Keynes MK1 1SW UK Phone 44 0 1908 274545 Fax 44 0 1908 641339 e mail sales adamequipment co uk...

Отзывы: