Setting of the check weighing alarm  

 

The display will show “

” where “ ” is a digit from   to  . 

 

  will set the buzzer to off when check weighing 

  will set the buzzer to on when the 

OK

 symbol is on 

  will set the buzzer to on when the weight is outside the 

OK

 symbol limit 

 

To change this digit press the

 [Tare/ ]

 key. 

 

Press the 

[Zero/Enter]

 key to store the changed value and continue to the next 

parameter. 

 

Speed at which the machine will run the ADC 

 

The slowest setting is 

 and the fastest is 

. The default setting is usually  .  

 
The display will show “

”. 

 

To change this setting press the

 [Tare/ ]

 key. 

 

Press 

[Zero/Enter]

 to store the changed value and return to the normal weighing. 

 

6.0  TWO POINT FACTORY CALIBRATION  

 

This procedure is the same as normal calibration but can be set even if the stability 

indicator will not show or when the raw counts do not seem to be moving. 

 

Please  ensure  that  the  load  cell  is  not  damaged  and  that  the  load  cell  is  wired 

correctly with the right output voltages. 

 

To start the factory calibration, turn the scale off and then turn it back on.  Press the 

[Limit]

 and 

[%]

 keys together during the initial counting from 9 to 0 on the display. 

 
The display will show "

".  Remove all weight from the pan and then press the 

[Zero/Enter]

 key to set the initial zero point when the scale is stable. 

 

The display will then show the first calibration weight used.  This value will be one 

third of the scale capacity.   

 

Place this value weight on the scale and press the 

[Zero/Enter]

 key when the stable 

indicator is on 

 

The  display  will  then  show  then  second  value  weight  to  be  used  which  will  be  full 

capacity of the scale. 

 

Place this value calibration weight on the platform and then press the 

[Zero/Enter]

 

key when the stable indicator is on.   

 

Remove the calibration weight as the scale counts back from 9 to 0. 

Содержание CBW 100a

Страница 1: ...Adam Equipment Company 2005 Adam Equipment CBW SERIES SERVICE MANUAL ADAM EQUIPMENT CO LTD p n 4331 Rev A3 January 2005...

Страница 2: ...0 ERROR CODES 5 4 0 CBW DESCRIPTION 6 5 0 CBW CALIBRATION PARAMETERS VER 1 2 2 03 7 5 1 CBW CALIBRATION VER 1 2 2 03 7 5 2 CBW PARAMETERS VER 1 2 2 03 8 6 0 TWO POINT FACTORY CALIBRATION 10 6 1 ADC OU...

Страница 3: ...is blank Power Switch faulty or not connected No turn on test Battery not charged or faulty Power supply faulty or incorrect type E6 is displayed Pan not installed Poor connection Load cell damaged El...

Страница 4: ...te weight Balance not level Obstruction between sample pan Wrong unit of weight displayed Cannot use Full Capacity Over load Stops hitting pan support or hitting bottom of load cell Electronics faulty...

Страница 5: ...Keyboard error Improper operation of the scale A D count is not correct when turning the scale on Platform is not installed Load cell is damaged Electronics is damaged Scale is not stable when settin...

Страница 6: ...W are similar except the selection of load cells and the program The main circuit board is the same as used in the CBC scales Normally if a problem is found with a circuit board the most cost effectiv...

Страница 7: ...ero Enter key when the stable indicator is on Or if a different value is desired enter the value using the arrow key to select the flashing digit and the to increment the flashing digit If value is co...

Страница 8: ...Print keys at the same time Setting the scale increment The display will show where can be 1 2 5 10 or 20 The first value shown is the default scale increment value To change to a different increment...

Страница 9: ...division The display will show Press the Tare key to change the value Press the Zero Enter key to store the changed setting Selectable unit of weight displayed The scales are normally set to display...

Страница 10: ...cator will not show or when the raw counts do not seem to be moving Please ensure that the load cell is not damaged and that the load cell is wired correctly with the right output voltages To start th...

Страница 11: ...ead load percentage of the cell If with no load on the pan the counts are below 1000 or above 200 000 then the load cell should be tested as explained in the next section Apply full capacity to the pa...

Страница 12: ...articular way on the metal structure These 4 strain gauges are connected in such a way that they form a wheatstone bridge See figure below In the simplest load cells there are only the 4 resistances o...

Страница 13: ...voltmeter with up to a 10volt power supply Before the load cell is connected to the power supply use the ohmmeter to measure the resistance between the wires Typical resistance values are E to E 410...

Страница 14: ...needs to be amplified to make it acceptable for the A D converter used The TCC CBW scales amplify the input by 200 and change the no load voltage to about 2 5V The TFC and TFW scales use a gain of 500...

Страница 15: ...he top pan and take out the 5 screws from the bottom of the scale 2 Detach the 4 cables connected with the Load cell Display Power Switch and Battery Chamber from the main PCB Remove the main PCB by t...

Страница 16: ...ale on to verify proper installation Set the parameters as described in the Service Manual and perform the Factory Calibration by using the Limit and keys together If the RS 232 needs to be disabled o...

Страница 17: ...Adam Equipment Company 2005 17...

Страница 18: ...Adam Equipment Company 2005 18...

Страница 19: ...to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment The equipment generates uses and can radiate radio frequency energy and if not...

Страница 20: ...tronic Scales for Medical use Retail Scales for price computing Adam Equipment Co Ltd Bond Avenue Milton Keynes MK1 1SW UK Phone 44 0 1908 274545 Fax 44 0 1908 641339 e mail sales adamequipment co uk...

Отзывы: