Calibration weights 

 

CBW SERIES 

 

CBW 3 

CBW 6 

CBW 15 

CBW 30 

CBW 45 

Weight 1 

1000 g 

2 kg 

5 kg 

10 kg 

15 kg 

Weight 2 

3000 g 

6 kg 

15 kg 

30 kg 

45 kg 

 

CBWa SERIES  

 

CBW 6a 

CBW 12a 

CBW 35a 

CBW 65a 

CBW 100a 

 

lb 

kg 

lb 

kg 

lb 

kg 

lb 

kg 

lb 

kg 

Weight 1 

2 lb 

1000 g 

4 lb 

2 kg 

10 lb 

5 kg 

20 lb 

10 kg 

40 lb 

15 kg 

Weight 2 

6 lb 

3000 g 

12 lb 

6 kg 

30 lb 

15 kg 

60 lb 

30kg 

100 lb 

45kg 

 

6.1   ADC OUTPUT CHECKING 

 

It is possible to check the raw count output of the 

A

nalogue to 

D

igital 

C

onverter 

which will also verify if the load cell is in good condition. 

 

To  start  the  ADC  check,  turn  the  scale  off  and  then  turn  it  back  on.    Press  the 

[Zero/Enter]

  and 

[%]

  keys  together  during  the  initial  counting  from  9  to  0  on  the 

display. 

 

The  raw  ADC  counts  will  be  displayed.  With  no  load  on  the  pan  the  raw  counts 

should typically be between 10,000 and 100,000 although this is not critical. 

 

Different capacity scales have different load cells and the pan assembly will vary the 

dead  load  percentage  of  the  cell.  If  with  no  load  on  the  pan  the  counts  are  below 

1000 or above 200,000 then the load cell should be tested as explained in the next 

section. 

 

Apply full capacity to the pan to verify that the raw counts increase to a value that will 

not  exceed  the  capability  of  the  ADC.  Typically  the  raw  counts  will  increase  to  a 

figure below 500,000 although if a mass has been added to the pan then this may be 

exceeded. If the counts go above 900,000 or decrease with load then please test the 

load cell for damage or incorrect wiring. 

 

To check if the ADC has  a good output for best accuracy then empty the pan  and 

press the 

[Zero/Enter]

 

 

key to set the counts to “0”. 

 

Now apply full capacity to the pan and note the span of raw counts.  

 

Divide  the  number  of  counts  by  the  number  of  displayed  divisions  (i.e.  30,000  as 

standard). If the input to the ADC is good then there should be approximately 10 or 

more counts per division for best accuracy. 

 

If there are less than 10 counts / d then check that the correct load cell is fitted and 

that the over load stops are adjusted correctly. 

 

To exit the ADC raw counts mode remove any weights and press the 

[Tare/ ] 

key. 

 

 

Содержание CBW 100a

Страница 1: ...Adam Equipment Company 2005 Adam Equipment CBW SERIES SERVICE MANUAL ADAM EQUIPMENT CO LTD p n 4331 Rev A3 January 2005...

Страница 2: ...0 ERROR CODES 5 4 0 CBW DESCRIPTION 6 5 0 CBW CALIBRATION PARAMETERS VER 1 2 2 03 7 5 1 CBW CALIBRATION VER 1 2 2 03 7 5 2 CBW PARAMETERS VER 1 2 2 03 8 6 0 TWO POINT FACTORY CALIBRATION 10 6 1 ADC OU...

Страница 3: ...is blank Power Switch faulty or not connected No turn on test Battery not charged or faulty Power supply faulty or incorrect type E6 is displayed Pan not installed Poor connection Load cell damaged El...

Страница 4: ...te weight Balance not level Obstruction between sample pan Wrong unit of weight displayed Cannot use Full Capacity Over load Stops hitting pan support or hitting bottom of load cell Electronics faulty...

Страница 5: ...Keyboard error Improper operation of the scale A D count is not correct when turning the scale on Platform is not installed Load cell is damaged Electronics is damaged Scale is not stable when settin...

Страница 6: ...W are similar except the selection of load cells and the program The main circuit board is the same as used in the CBC scales Normally if a problem is found with a circuit board the most cost effectiv...

Страница 7: ...ero Enter key when the stable indicator is on Or if a different value is desired enter the value using the arrow key to select the flashing digit and the to increment the flashing digit If value is co...

Страница 8: ...Print keys at the same time Setting the scale increment The display will show where can be 1 2 5 10 or 20 The first value shown is the default scale increment value To change to a different increment...

Страница 9: ...division The display will show Press the Tare key to change the value Press the Zero Enter key to store the changed setting Selectable unit of weight displayed The scales are normally set to display...

Страница 10: ...cator will not show or when the raw counts do not seem to be moving Please ensure that the load cell is not damaged and that the load cell is wired correctly with the right output voltages To start th...

Страница 11: ...ead load percentage of the cell If with no load on the pan the counts are below 1000 or above 200 000 then the load cell should be tested as explained in the next section Apply full capacity to the pa...

Страница 12: ...articular way on the metal structure These 4 strain gauges are connected in such a way that they form a wheatstone bridge See figure below In the simplest load cells there are only the 4 resistances o...

Страница 13: ...voltmeter with up to a 10volt power supply Before the load cell is connected to the power supply use the ohmmeter to measure the resistance between the wires Typical resistance values are E to E 410...

Страница 14: ...needs to be amplified to make it acceptable for the A D converter used The TCC CBW scales amplify the input by 200 and change the no load voltage to about 2 5V The TFC and TFW scales use a gain of 500...

Страница 15: ...he top pan and take out the 5 screws from the bottom of the scale 2 Detach the 4 cables connected with the Load cell Display Power Switch and Battery Chamber from the main PCB Remove the main PCB by t...

Страница 16: ...ale on to verify proper installation Set the parameters as described in the Service Manual and perform the Factory Calibration by using the Limit and keys together If the RS 232 needs to be disabled o...

Страница 17: ...Adam Equipment Company 2005 17...

Страница 18: ...Adam Equipment Company 2005 18...

Страница 19: ...to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment The equipment generates uses and can radiate radio frequency energy and if not...

Страница 20: ...tronic Scales for Medical use Retail Scales for price computing Adam Equipment Co Ltd Bond Avenue Milton Keynes MK1 1SW UK Phone 44 0 1908 274545 Fax 44 0 1908 641339 e mail sales adamequipment co uk...

Отзывы: