© Adam Equipment Company 2007
5
Off Centre Loading error
Overload Stops not correct
Obstruction
under
pan
Battery will not charge
Charging circuit failure
Battery Failure
Incorrect adapter (below 800 mA)
Processor failure
4.0 ERROR
CODES
ERROR
CODE
DESCRIPTION POSSIBLE
CAUSES
Err 1*
Time input Error
Invalid time entry such as “268970” for
the time format “
H-m-S
”.
Err 2*
Date input Error
34
th
day of a month is an invalid entry.
Err 4
Initial Zero is greater than
allowed (4% of maximum
capacity) when power is turned
on or when the
[Zero/Enter]
key is pressed.
Weight on the pan when turning the
scale on.
Excessive weight on the pan when
zeroing the scale.
Platform is not installed.
Improper calibration of the scale.
Damaged load cell.
Damaged Electronics.
Err 6
A/D count is not correct when
turning the scale on.
Load cell is damaged.
Electronics is damaged.
Err 7
Percent
input
error
Percent function is entered with no
reference mass on the pan.
Err 8
High limit input error
Low limit is set first, then the high limit is
set lower than the low limit and high
limit not equal to zero.
Err 9
Low limit input error
High limit is set first, then the low limit is
set higher than the high limit and low
limit not equal to zero.
FAIL H
or
FAIL L
Calibration error
Improper factory calibration). The old
calibration data will be retained until the
calibration process is complete.
*Err 1
and
Err 2
are not applicable for scales used in approved applications.