8.0 FACTORY
PARAMETERS
The factory parameters are used performing the factory calibration, setting the capacity
and other basic settings.
To access these parameters place a jumper into the pins on the main PCB (marked as
position J4) through a hole provided on the base of the scale.
Turn the scale upside down.
Remove the seal covering the hole to access the PCB.
Place a jumper through the hole into the pins. If a
jumper is not available, short the pins.
Turn on the scale using the switch on the base.
“
P _ _ _ _
”
Enter
“2006” and p
ress
[Tare]
After the jumper has been placed in the position shown, apply power to the scale.
The display will show the software revision number and then ask for a code number,
“
P _ _ _ _
“.
Enter the number 2006 and press
[Tare].
The display will show the first parameter
“
F1 CAL
”.
To scroll through other parameters
press the
[Func]
key.
© Adam Equipment Company 2007
11