@Adam Equipment Company 2005
8
5.0 CBC PARAMETERS AND CALIBRATION (VER. 1.2- 2.02)
5.1 CBC PARAMETERS (VER. 1.2- 2.02)
To set the parameters it is necessary to enter a secure menu by entering a password number
when requested.
To enter the parameter menus press
[Tare]
once, during the initial counting of the display
after the power is turned on.
The “
Weight
” display will show "
" requesting the password number to be entered. The
default password is "
" but other numbers can be set using the parameter menus. Press
the
[0]
key four times.
Press the
[Tare]
key.
The Parameter menu has 6 functions that can be accessed using the
[U. Wt./Units]
key to
cycle through the choices. The “
Weight
” display will show the name of the functions. To
enter a function, press
[Tare]
.
At any time press the
[Zero]
key to return to weighing.
Weight Display Description
See the Calibration section for details.
Sets the scale increment. Press the
[U. Wt./Units]
key to cycle
through the options (1,2,5,10). Press
[Tare]
to accept the selection.
Displays the A/D counts. Press
[Tare]
to return to menu.
Press the
[U. Wt./Units]
key to show the options. Select automatic
accumulation (
) when the scale becomes stable, Manual
accumulation (
) or Continuous (
), the user must press the
[M+]
key to accumulate data.
Press
[U. Wt./Units]
to show the options for the auto-zero range.
The value can be set to 0.5d, 1d, 2d or 4d. This is normally set to 1d
but can be increased to force the scale to zero if it is likely to have
small amounts of material spilled on the platform while in use.
Set a new password number. Display will show "
"
Enter the new password number then press the
[Tare]
key.
Display will change to "
", Enter the password again and press
[Tare]
again.
The display will show "
" to show the new password has been
accepted.
Record the new password number in a secure place.
This is used to set the speed at which the scale will run the ADC. The
settings are 7.5, 15, 30 and 60. The slowest setting is 7.5 and the
fastest is 60.