Adam Equipment CBC 15 Скачать руководство пользователя страница 11

@Adam Equipment Company 2005

 

 

 

11 

Place this weight on the scale and press the 

[Tare]

 key after the stable symbol is on.   

 

The display will now show the second weight needed.  This will be full capacity. 

 

Place the weight on the platform and press the 

[Tare]

 key again after the stable symbol is on. 

 

When the scale begins its start up procedure the weights can be removed and calibration is 

complete. 

 

Suggested Factory Calibration weights: 

 

CBC SERIES 

 

CBC 3 

CBC 6 

CBC 15 

CBC 30 

CBC 45 

Weight 1 

1000 g 

2 kg 

5 kg 

10 kg 

15 lb 

Weight 2 

3000 g 

6 kg 

15 kg 

30 kg 

45 kg 

 

CBCa SERIES  

 

CBC 6a 

CBC 12a 

CBC 35a 

CBC 65a 

CBC 100a 

 

lb 

kg 

lb 

kg 

lb 

kg 

lb 

kg 

lb 

kg 

Weight 1 

2 lb 

1000g 

4 lb 

2 kg 

10 lb 

5 kg 

20 lb  10 kg  40 lb. 

15 kg 

Weight 2 

6 lb 

3000g 

12lb 

6 kg 

30 lb 

15 kg 

60 lb  30 kg  100 lb  30 kg 

 

8.0  LOAD CELL DAMAGE 

 

The most common reason a scale fails is that the load cell has been damaged.  The damage 

can be from 2 primary causes.  The first is physical damage due to an overload or an impact 

from the side and the second is damage due to the environment, such as moisture, extreme 

heat or a cut cable if the cables are exposed. 

 

 

PRINCIPLE OF OPERATION 

 

The stain  gauge load  cell  is  a method  of measuring the  amount  of stress  put  onto  a  metal 

structure due to a weight being supported by the structure.  The stress is measured using 4 

strain gauge sensors mounted in a particular way on the metal structure.   

 

These 4 strain gauges are connected in such a way that they form a wheatstone bridge.  See 

figure below.  In the simplest load cells there are only the 4 resistances of the strain gauges to 

consider.  Normally without a load on the load cell all the resistances are the same.  However 

when the load cell has a weight on it 2 of the strain gauges will be in compression (A and D) 

and the resistance will decrease and 2 will be in tension (B and C) and their resistance will 

increase. 

Содержание CBC 15

Страница 1: ...Adam Equipment Company 2005 Adam Equipment CBC SERIES SERVICE MANUAL ADAM EQUIPMENT CO LTD p n 4329 Rev A3 January 2005...

Страница 2: ...Adam Equipment Company 2005 2...

Страница 3: ...6 4 0 CBC DESCRIPTION 7 5 0 CBC PARAMETERS AND CALIBRATION VER 1 2 2 02 8 5 1 CBC PARAMETERS VER 1 2 2 02 8 5 2 CBC CALIBRATION VER 1 2 2 02 9 6 0 FACTORY PARAMETERS VERSION 1 1 1 8 TO 1 2 2 02 10 7 0...

Страница 4: ...your dealer or Adam Equipment for more assistance In order to supply further assistance we will need the following information Name of your company Contact Name Contact telephone e mail fax or other m...

Страница 5: ...ng animal weighing Vibrations through table Temperature changed dramatically Electronics faulty Battery low Weight value incorrect Calibration error Recalibrate Unit calibrated with inaccurate weight...

Страница 6: ...alibration of the scale Damaged load cell Damaged Electronics Keyboard error Improper operation of the scale A D count is not correct when turning the scale on Platform is not installed Load cell is d...

Страница 7: ...C are similar except the selection of load cells and the program The main circuit board is the same as used in the CBW scales Normally if a problem is found with a circuit board the most cost effectiv...

Страница 8: ...the scale increment Press the U Wt Units key to cycle through the options 1 2 5 10 Press Tare to accept the selection Displays the A D counts Press Tare to return to menu Press the U Wt Units key to...

Страница 9: ...t on the platform after the stable symbol is on or enter the value desired and then press Tare The display will show Place the weight on the platform and press Tare again At all times the scale should...

Страница 10: ...nting of the display after power turned on The Weight Display will show requesting the password number to be entered The factory password is and will over ride any other password they may have been pr...

Страница 11: ...is that the load cell has been damaged The damage can be from 2 primary causes The first is physical damage due to an overload or an impact from the side and the second is damage due to the environmen...

Страница 12: ...al output at S and S will show a voltage of zero volts with no load and a small voltage proportional to the load as the load is increased Typical load cells show 20 30mv of signal if the excitation vo...

Страница 13: ...or 3mv V The expected change to the output is Vout 2mv V Vext Mass capacity of load cell For example a 2mv V load cell of 30Kg capacity with 20Kg placed on it Vext is 5 volts would give a output of a...

Страница 14: ...arameters and Calibration including the Two Point Factory Calibration must be set before the new PCB will be put to use for weighing The 3 pin connector from the RS 232 connecter goes to the main PCB...

Страница 15: ...processor and the links in the same way as it was in the old board into the new board Be very careful not to bend the legs while removing the micro processor Plug the cables coming from the Load cell...

Страница 16: ...t advice you to change parts on the circuit boards or otherwise disturb the electronics If you require assistance please contact your dealer or Adam Equipment We will be happy to help you resolve prob...

Страница 17: ...Adam Equipment Company 2005 17...

Страница 18: ...Adam Equipment Company 2005 18...

Страница 19: ...to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment The equipment generates uses and can radiate radio frequency energy and if not...

Страница 20: ...tronic Scales for Medical use Retail Scales for price computing Adam Equipment Co Ltd Bond Avenue Milton Keynes MK1 1SW UK Phone 44 0 1908 274545 Fax 44 0 1908 641339 e mail sales adamequipment co uk...

Отзывы: