Adam Equipment CBC 15 Скачать руководство пользователя страница 13

@Adam Equipment Company 2005

 

 

 

13 

Before  the  load  cell  is  connected  to  the  power  supply  use  the  ohmmeter  to  measure  the 

resistance between the wires.  Typical resistance values are: 

 

 

+E to -E 

 

410 ohms  ±30 ohms 

 

+S to -S 

 

350 ohms  ±  2 ohms 

 

+E to +S or any other combination similar approximately 270 ohms. 

 

If the load cell has Sense connections in addition they are connected to the excitation internal 

to the load cell. 

 

Connect the load cell +E and -E to a power supply, typically 10 volts.  Never use more than 

12 volts as it may damage the load cell. Measure the voltage bS and -S.   

 

With no load the voltage should be approximate 0mv ±3mv. 

 

With a mass on the load cell the output voltage should increase.  The amount of increase will 

depend upon the sensitivity  of the load cell, capacity of the load  cell, the excitation voltage 

and the amount of mass placed on the load cell.  Most load cells have a sensitivity of either 

2mv/V or 3mv/V. 

 

The expected change to the output is: 

 

 

Vout =  (2mv/V) * Vext * Mass 

 

 

  capacity of load cell 

 

For example a 2mv/V load cell of 30Kg capacity with 20Kg placed on it.  Vext is 5 volts would 

give a output of approximately: 

 

 

Vout =  (2mv/V) * Vext * Mass      = 2mv/V * 5V  *  20Kg   = 6.7mv 

 

 

  capacity of load cell 

 

   30kg 

 

If the load cell has been damaged the no load voltage will likely be greater than 3mv or the 

loaded  voltage  will  be  grossly  different  from  the  expected  value.    Remember  the  loaded 

voltage will be offset by the amount of zero load voltage measured. 

 

The load cell can be tested while it is connected to the A/D converter circuit board, using the 

scale power supply for excitation voltage.   

 

 

TESTING LOAD CELLS IN THE SCALE 

 

The signal from the load cells is amplified by the circuits on the A/D circuit board.   

 

The  signal  needs  to  be  amplified  to  make  it  acceptable  for  the  A/D  converter  used.    The 

CBC/CBW scales amplify the input by 200, and change the no load voltage to about +2.5V.     

 

The amplified and offset voltage is measured at pin 6 of IC7, the INA118P amplifier IC.  This 

voltage is sent to the filter and then to the A/D converter. 

Содержание CBC 15

Страница 1: ...Adam Equipment Company 2005 Adam Equipment CBC SERIES SERVICE MANUAL ADAM EQUIPMENT CO LTD p n 4329 Rev A3 January 2005...

Страница 2: ...Adam Equipment Company 2005 2...

Страница 3: ...6 4 0 CBC DESCRIPTION 7 5 0 CBC PARAMETERS AND CALIBRATION VER 1 2 2 02 8 5 1 CBC PARAMETERS VER 1 2 2 02 8 5 2 CBC CALIBRATION VER 1 2 2 02 9 6 0 FACTORY PARAMETERS VERSION 1 1 1 8 TO 1 2 2 02 10 7 0...

Страница 4: ...your dealer or Adam Equipment for more assistance In order to supply further assistance we will need the following information Name of your company Contact Name Contact telephone e mail fax or other m...

Страница 5: ...ng animal weighing Vibrations through table Temperature changed dramatically Electronics faulty Battery low Weight value incorrect Calibration error Recalibrate Unit calibrated with inaccurate weight...

Страница 6: ...alibration of the scale Damaged load cell Damaged Electronics Keyboard error Improper operation of the scale A D count is not correct when turning the scale on Platform is not installed Load cell is d...

Страница 7: ...C are similar except the selection of load cells and the program The main circuit board is the same as used in the CBW scales Normally if a problem is found with a circuit board the most cost effectiv...

Страница 8: ...the scale increment Press the U Wt Units key to cycle through the options 1 2 5 10 Press Tare to accept the selection Displays the A D counts Press Tare to return to menu Press the U Wt Units key to...

Страница 9: ...t on the platform after the stable symbol is on or enter the value desired and then press Tare The display will show Place the weight on the platform and press Tare again At all times the scale should...

Страница 10: ...nting of the display after power turned on The Weight Display will show requesting the password number to be entered The factory password is and will over ride any other password they may have been pr...

Страница 11: ...is that the load cell has been damaged The damage can be from 2 primary causes The first is physical damage due to an overload or an impact from the side and the second is damage due to the environmen...

Страница 12: ...al output at S and S will show a voltage of zero volts with no load and a small voltage proportional to the load as the load is increased Typical load cells show 20 30mv of signal if the excitation vo...

Страница 13: ...or 3mv V The expected change to the output is Vout 2mv V Vext Mass capacity of load cell For example a 2mv V load cell of 30Kg capacity with 20Kg placed on it Vext is 5 volts would give a output of a...

Страница 14: ...arameters and Calibration including the Two Point Factory Calibration must be set before the new PCB will be put to use for weighing The 3 pin connector from the RS 232 connecter goes to the main PCB...

Страница 15: ...processor and the links in the same way as it was in the old board into the new board Be very careful not to bend the legs while removing the micro processor Plug the cables coming from the Load cell...

Страница 16: ...t advice you to change parts on the circuit boards or otherwise disturb the electronics If you require assistance please contact your dealer or Adam Equipment We will be happy to help you resolve prob...

Страница 17: ...Adam Equipment Company 2005 17...

Страница 18: ...Adam Equipment Company 2005 18...

Страница 19: ...to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment The equipment generates uses and can radiate radio frequency energy and if not...

Страница 20: ...tronic Scales for Medical use Retail Scales for price computing Adam Equipment Co Ltd Bond Avenue Milton Keynes MK1 1SW UK Phone 44 0 1908 274545 Fax 44 0 1908 641339 e mail sales adamequipment co uk...

Отзывы: