4.0 ERROR CODES
During the initial power-on testing or during operation, the scale may show an
error message. If an error message is shown, repeat the step that caused
the message. If the error is still shown contact your dealer for further support.
ERROR DESCRIPTION
POSSIBLE CAUSES
Err 1
Time input error.
Tried to set an illegal time, i.e.
26 hours
Err 2
Date input error
Tried to set an illegal date, i.e.
36
th
day
Err 4
Initial Zero is greater than
allowed (typically 4% of
the maximum capacity)
when power is turned on
or when [Zero] is
pressed,
Weight is on the pan when
turning the scale on.
Excessive weight on the pan
when zeroing scale.
Improper calibration of the
scale.
Damaged load cell.
Damaged Electronics.
Err 6
A/D count is not correct
when turning the scale
on.
Platform is not installed.
Load cell is damaged.
Electronics is damaged.