© Adam Equipment Company 2000 

10 

 

8.0 PARAMETER SET-UP 

 

 

The parameter set-up menus allow the scale to be set as required by the user. It is 

possible to calibrate the scale, enable specific units of weight (up to 13 units), select the 

default unit of weight, control autozero, the RS-232 interface, digital filtering and set the 

automatic turn-off time. The parameters are selected using the function menu. An 

overview of the function menu is shown as Appendix A. The specific steps to set the 

functions are described below. 

 

To start the function menu: 

 

1. 

Press and hold the MODE key, meanwhile also press the ENTER/select key for 2 

seconds then release the keys. The display will show CAL and enter to the 

FUNCTION SELECTIONS. 

 

2. 

Press the PCS/next key to choose the function desired. 

 

3. 

The functions will change each time the PCS/next key is pressed. The functions 

are: 

CAL – InitU – Unit – A.Off – trn – rAtE – Filxx – Zerox – b2x – Acy – trACx – CAL- - 

 

 

8.1  “CAL”  Auto Calibration, Customised weight. 

 

Select the CAL function by pressing the ENTER/select key. 

 

Press the LIGHT/increment key and ENTER/select key to key in the mass weight which 

you are going to calibrate (preferably 2/3 to full capacity).  The default weight value is 

initially shown.   

On the last press of ENTER/select the display will show the  weight entered with the 

digits flashing.   Now put this mass on the pan. When stable, the display will become 

stable with the digits not flashing.  The calibration is finished. 

 

 

8.2  “InitU”  To choose the initial unit while turning on: 

 

Press the ENTER/select key to set the initial units. 

 

Please press the UNITS key to choose the unit. Then press the ENTER/select key to 

confirm and return to normal weighing or press the PCS/next key to continue to another 

function mode. 

 

 

Содержание ACB-1200H

Страница 1: ...Adam Equipment Company 2002 Adam Equipment ACB SERIES ADAM EQUIPMENT CO LTD pn 3426 rev H November 02...

Страница 2: ...to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment The equipment generates uses and can radiate radio frequency energy and if not...

Страница 3: ...CAL Auto Calibration Customised weight 10 8 2 InitU To choose the initial unit while turning on 10 8 3 Unit To choose the weighing units to enable 11 8 4 A OFF To choose the time of AUTO OFF 11 8 5 tr...

Страница 4: ...Adam Equipment Company 2000 1...

Страница 5: ...ud 1 1 INSTALLING THE BALANCE Unpack the balance carefully Place the scale on a stable surface The weighing area should be kept clean and dry Do not install the balance near extremes of heat and cold...

Страница 6: ...The 4 sides slide together to make a box then the top can be placed over the top for additional protection The breeze shield fits on top of the scale positioned by the ridges on the cover 1 3 DISPLAY...

Страница 7: ...ng counting or percentage weighing mode Select Parameter Set up Mode UNITS unit Select weighing units Select weighing units to set as initial unit or enable PCS next Set initial sample size during par...

Страница 8: ...tracted by placing an empty container on the scale When the display is stable press the TARE key The display will become zero and the display will have a tare indication Cancel the tare function by pr...

Страница 9: ...eighing with the MODE key press the UNITS key to choose either 100 0 100 00 b Put the item to be considered 100 onto the pan and press the PCS key During the sampling the symbol will be flashing and w...

Страница 10: ...he display will show the Calibration Weight Value Put the same weight on the platform When the display is stable the scale will calibrate Verify the calibration by removing the mass and checking it ag...

Страница 11: ...ed from the scale Do not use the power supply at the same time as batteries are installed Remove the batteries first Power supplies 1 9V 300mA AC DC adapter 2 4 pcs C size dry cell batteries Batteries...

Страница 12: ...ted it will print once automatically when the scale becomes stable and if ser has been selected it will print continuously The data format is as follows 20 characters are sent with the following field...

Страница 13: ...desired 3 The functions will change each time the PCS next key is pressed The functions are CAL InitU Unit A Off trn rAtE Filxx Zerox b2x Acy trACx CAL 8 1 CAL Auto Calibration Customised weight Selec...

Страница 14: ...ct no to disable auto off Then press ENTER select to return to weighing or PCS next key to continue to another function mode 8 5 trn To choose RS 232 TRANSMISSION METHOD Press the ENTER select key to...

Страница 15: ...to confirm and return to normal weighing or press the PCS key to continue to another function mode 8 9 b2 x To choose the Auto Zero recovery range Press the LIGHT increment key to choose the Auto Zero...

Страница 16: ...to choose the range from 0 to 5 Then press the ENTER key to confirm and return to normal weighing or press the PCS key to continue to another function mode Note When the scale is in the function mode...

Страница 17: ...0 0005 42 x 0 001 dr 84 x 0 005 160 x 0 01 330 x 0 02 840 x 0 05 1600 x 0 1 330 x 0 01 660 x 0 02 GN 2300 x 0 1 4600 x 0 2 9200 x 0 5 23000 x 1 46000 x 2 9200 x 0 2 18000 x 0 5 ozt 4 8 x 0 0002 9 6 x...

Страница 18: ...POUND 453 59237 g 1 oz AVOIRDUPOIS OUNCE 28 349523 g 1 dr AVOIRDUPOIS DRAM 1 7718451 g 1 gn GRAIN UK 0 0647989 g 1 ozt TROY OUNCE 31 103476 g 1 dwt PENNYWEIGHT 1 555174 g 1 MM MOMME Japan 3 749996 g...

Страница 19: ...Adam Equipment Company 2000 16 APPENDIX PARAMETER SETTING MENU...

Страница 20: ...laboratory balances Basic range of top loading balances for schools and educational establishments Counting scales for general industrial and warehouse applications Digital weighing check weighing sca...

Отзывы: