3. Apply control solution
Shake the control solution vial thoroughly before use.
Squeeze out a drop and wipe it off, then squeeze out
another drop and place it on the tip of the vial cap.
Hold the meter to move the absorbent hole of the test strip
to touch the drop. Once the confirmation window fills
completely, the meter will begin counting down.
To avoid contaminating the control solution, do not directly apply control solution
onto a strip.
4. Read and compare the result
The control solution test result will appear on the display in seconds. Compare this
result with the range printed on the test strip vial and it should fall within this range.
If not, please read the instructions again and repeat the control solution test.
Out-of-range Results
If you continue to have test results fall outside the range printed on the test strip
vial, the meter and strips may not be working properly. Do NOT test your blood.
Contact the service center for help.
NOTE:
The control solution range printed on the test strip vial is for control solution use
only. It is not a recommended range for your blood glucose level.
See the
Maintenance
section for important information about your control
solutions.