Ultrasonic tester UK1401
Operation Manual
33
ACOUSTIC
CONTROL
SYSTEMS
2
2.3.2.2 RECORDING the Result
The measurement results are stored in the cells of the memory. The cells form groups. The groups and the cells in the
groups are identified by the serial numbers. The numbering of the groups and the cells in each group is started with “1”.
The maximum quantity of the cells in a group is 500.
The maximum quantity of all groups is 100.
When the quantity of the cells in a group becomes maximum, a corresponding message is indicated on the display.
In practice, it is more convenient to record the results into small groups (by several dozen values). You can return to any
existing group if required and continue recording the results in it.
To record a result, select any existing group or create a new one prior to measurements, and press the ENTER
button. The measurement result will be saved in the last cell of the current group.
2.3.2.3 Viewing and adjusting the measurement results
The user can view any result recorded in the memory. The result can be measured again in the same point and
overwritten if the user doubts its reliability.
To see the results, switch to the CELL column. The character will be displayed on the screen (Figure 34).
Scrolling along the measurement results is performed successively by the cells of the groups according to the
selected view direction. When the first/last cell in the group is reached, the program will switch to the next/previous
group of the results correspondingly.
To correct a result, please perform the following:
- using the keys switch to the cell with the recorded result which will be corrected;
- while pressing and holding the
key, switch to the RESULT column, at that the
sign will dissapear.
Note:
Actually, the result will remain in the memory up to the moment it will be recorded to the selected cell of
new purpose. To return to the view mode not changing the result in the cell, press the
key.
- perform the measurement. If the measurement result is satisfactory, then press the “ENTER” button to record it
into the selected cell. The instrument will automatically return to the view mode after doing it.
Figure 34
Содержание UK1401
Страница 2: ...Ultrasonic tester UK1401 Operation Manual 2 ACOUSTIC CONTROL SYSTEMS...
Страница 41: ...Ultrasonic tester UK1401 Operation Manual 41 ACOUSTIC CONTROL SYSTEMS NOTES...
Страница 42: ...Ultrasonic tester UK1401 Operation Manual 42 ACOUSTIC CONTROL SYSTEMS NOTES...
Страница 43: ...Ultrasonic tester UK1401 Operation Manual 43 ACOUSTIC CONTROL SYSTEMS...