Ultrasonic tester UK1401
Operation Manual
19
ACOUSTIC
CONTROL
SYSTEMS
2
Table 4
Menu option
(parameter)
Parameter value
Description
Mode
Time / Velocity /
Crack / Front
Selects the measurement mode
Accumulation, Qty
1 / 2 / 4 / 8 / 16 / 32 / 64 Sets quantity of accumulations
AGC
On / Off
Switches On/Off the automatic control system of the receive path gain
Gain, dB
(only for AGC Off)
from 40 to 80
in increments of 2
Sets the gain
Frequency, Hz
from 2 to 25
Repetition frequency of the sounding pulses
Norm, µs
(in the FRONT mode only)
from 0.1 to 20.0
Sets the normative value of duration of the rising edge (front) of the
signal
Clear memory, %
Occupied memory
volume in %
Starts the memory cleaning procedure
Calibration
–
Starts the calibration procedures
Delay, µs
from 0 to 20
Instrument-specific parameter associated with constructive features
Base, mm
from 149 to 151
Actual value of the instrument's scanning base
Sound
On / Off
Monitoring the sound indication
Vibration
On / Off
Monitoring the vibration indication
Language
Russian / English
Selects the interface language
Measuring units
mm / inches
Selects the measurement unit system
Brightness, %
from 10 to 100
Sets the display brightness
Current time
Editable
Sets date and time
Содержание UK1401
Страница 2: ...Ultrasonic tester UK1401 Operation Manual 2 ACOUSTIC CONTROL SYSTEMS...
Страница 41: ...Ultrasonic tester UK1401 Operation Manual 41 ACOUSTIC CONTROL SYSTEMS NOTES...
Страница 42: ...Ultrasonic tester UK1401 Operation Manual 42 ACOUSTIC CONTROL SYSTEMS NOTES...
Страница 43: ...Ultrasonic tester UK1401 Operation Manual 43 ACOUSTIC CONTROL SYSTEMS...