1
А1020 MIRA Lite Low-Frequency Ultrasonic Tomograph
Operation Manual
18
ACOUSTIC
CONTROL
SYSTEMS
1.4.1.1 Editing parameters
All parameters of the SETUP mode can be configured.
The functions of the keys for configuring the parameters are listed in the Table 3.
Icon
Purpose
Scroll the rows to select a parameter to be configuredw
Change the parameter value
Exit the SETUP mode
Table 3
Table 4
The parameter description and their permissible values are listed in the Table 4.
Parameter
Value
Description
Color gain, dB
from 0 to 60 in increments of 1
Changes the brightness scale for the synthesized image
Analog gain, dB
from 0 to 100 in increments of 1
Selects the gain coefficient in the path of the device to pro-
vide maximum dynamic range without the signal limitation
Filter, kHz
10 / 20 / 40 /100
Sets the high-frequency filter