Model TT339-0700
4-Wire Isolated Frequency Transmitter w/USB
Acromag, Inc. Tel: 248-295-0880
[
22
]
http://www.acromag.com
Configuration...
HELP:
Press
[F1]
for Help on a selected or highlighted field
or control. Alternatively, click the
[?]
button in the
upper-right hand corner of the window and click a
field or control to receive a Help message pertaining
to the item pointed to.
Submit I/O Configurations
Once you have made your configuration selections, click the
[Send I/O Config]
button to write them to the transmitter. You
can read the status of your sent message in the “Status” field.
Alternately, you could click
“File”
in the upper left-hand corner
to save the configuration settings to a file on your PC for later
reference.
At this point, you can test the transmitter’s operation by clicking
on the
[Start Polling]
button of the TEST I/O Section of the page
to trigger the software to periodically read the input and display
its value in the field to the right of the polling toggle button.
Note the simulated lamp next to the button flashes each time
the input is sampled. Click
[Stop Polling]
to stop polling the
input.
Other Configuration Controls
Restore Factory Calibration:
Resets transmitter and causes it to
revert to its factory calibration without effecting user
configuration. Useful if there was an error during recalibration
that degraded performance or the I/O channel appears erratic.
Restore Factory Default:
Restores a transmitter to its original
factory state and configuration settings (See Specifications
Reference Test Conditions). This does not restore calibration,
only configuration. Alternately, this control can be used as a
sanitation tool to restore the transmitter to its initial
configuration.