Model TT339-0700
4-Wire Isolated Frequency Transmitter w/USB
Acromag, Inc. Tel: 248-295-0880
[
16
]
http://www.acromag.com
Quick Overview - Windows..
HELP:
You can press
[F1]
for Help on a selected
or highlighted field or control. You can
also click the
[?]
button in the upper-right
hand corner of the window and click to
point to a field or control to get a Help
message pertaining to the item you
pointed to.
Calibration (Calibrate the Output if Needed)
This transmitter has already been factory calibrated. If you encounter
excessive error, you can click the Calibration tab to display the Calibration
control page as shown at left.
For output zero-scale endpoint adjustment, click the
[Zero-Scale]
button
to open the zero-scale configuration popup. Use a current meter or digital
volt meter to accurately measure the transmitter output signal and enter
the measurement in the input field provided. Click the
[Submit
Measurement]
button to send the calibration to the transmitter.
For output full-scale endpoint adjustment, click the
[Full-Scale]
button to
open the full-scale configuration popup. Use a current meter or digital volt
meter to accurately measure the transmitter output signal and enter the
measurement in the input field provided. Click the
[Submit Measurement]
button to send the calibration to the transmitter.
Factory Settings (Use only in case of trouble or for sanitation purposes)
•
Restore a transmitter to its original factory calibration
•
Restore a transmitter to its initial factory configuration
You can click the
[Restore Factory]
buttons if you ever misconfigure or
miscalibrate a transmitter in such a way that its operation appears erratic.
Calibration Status (Bottom of Window)
The Calibration Status bar at the bottom of the window will display status
messages relative to calibration.