Step 4: Install the Card
Acqiris SA230P Startup Guide
15
SA230P Front Panel Features
Front Panel Connectors
Connector
Type
Description
TRG IN
MMCX
female
External trigger input, 50 Ω DC terminated, ± 5 V range.
IN 1, 2
SMA
female
Analog signal inputs, DC-coupled and 50 Ω terminated. The input full scale
ranges are selectable:
Voltage
500 mV FSR
2.5 V FSR
Recommended maximum
operating voltage
± 600 mVpk
±3 Vpk
TRG OUT
1
MMCX
Trigger Out signal (programmable).
50 Ω source, LVCMOS 3.3 V
I/O 1, 2, 3
MMCX
User configurable digital Input / Output signal.
DC coupling, LVCMOS 3.3 V.
Output: 50 Ω source, Input: +5 V max.
REF IN
MMCX
External reference clock input, AC coupled and 50 Ω terminated.
It can accept a 10 MHz or a 100 MHz signal from -3 to +3 dBm.
The ADC card can usually work with signal present at the external reference input (REF IN).
However, to ensure the best performance, or if the calibration is found to be unreliable, it is
recommended to remove such signals when working with internal clock.
1
The trigger out connector depends on the product version.