MEMBRANE SUCTION-IRRIGATION PUMP
Ackermann
Page: 8 // 20
7.2. REPROCESSING THE SPIkE
Read all instructions carefully before reprocessing the spike. For pre-cleaning/cleaning we advise to use Neodisher FA
(manufactured by Dr. Weigert). Pre-cleaning prevents the transmission of germs and stops residues from drying on. It
therefore needs to be performed immediately after the operation. Rinse spike below the surface of the water. Use cleaning
gun to rinse connectors and cannula at 4 bar / 3 min with cold water. Use a plastic brush to remove heavily Encrusted
residues.
WARNING!
Do not use any fixatives. Do not use hot water for pre-cleaning.
For manual cleaning procedure clean internal spaces with cleaning brush and use cleaning gun to rinse at 45-85 °C /
3-20 min. In addition you can clean the spike in an ultrasound bath (40°C; 35-45 kHz; min. 4 min). Turn the spike several
times. rinse with demineralized water, disinfect at 93-98°C and dry inside and out at 50-100°C/min.10 min. make sure
that venting hole is completely open.
CAUTION!
Do not use abrasive brushes or sponges. Avoid contact with hydrogen peroxide (H2O2).
use the sterilizer model Selectomat S 3000 made by mmm group or vario-calv 400 e made by Fisher Scientific. Sterilize
the spike at 134-137 °C, 3bar / 8-30 min. make sure that spike does not touch other components and the steam can
circulate freely.
8. SECRETION CONTAINER
8.1. HanDlIng oF SeCreTIon ConTaIner
The provided secretion container is to be used for collecting the aspired organic fluid and has a maximum capacity of 2
litres. It is equipped with pressure lid, mechanical overflow protection with integrated float and a volume scale.
Close the lid by pressing it onto the secretioncontainer. Ensure that integrated float of overflow protection can move
freely.
Connect tubing set SUCTION to the connectors „VACUUM“ and „PATIENT“ (see chapter 9.4.). Both connectors must not be
mixed up! Inversion of connections could lead to contamination of the operator and/or the pump.
Place secretion container with overflow protection only in upright position! Use bottle holder on pump.
When overflow valve intervenes, switch off pump immediately.
CAUTION!
Clean and sterilize secretion container and its components before initial application and any further use.
Operate secretion container in upright position only! Do not operate without overflow protection!
8.2. REPROCESSING OF SECRETION CONTAINER
Read all instructions carefully before reprocessing the secretion container. First of all disassemble secretion container
into its single components (see illustration).
Wash all surfaces using hot water (max. 60°C) and dry them carefully. All components are to sterilize by steam sterilisation
(saturated vapour) at 121°C / max.15min. Therefore place the container upside-down.
Carefully check components of secretion container for cracks and brittle parts after each cleaning cycle.
ATTENTION!
Do not use solvents, disinfectants or abrasive products for cleaning! Obey sterilization condition! Dispose
after 30 cleaning and sterilization cycles!
IMPORTANT!
Make sure to re-assemble the seals correctly and ensure that integrated float of overflow protection can
move freely after reprocessing.