MEMBRANE SUCTION-IRRIGATION PUMP
Ackermann
Page: 14 // 20
WARNING!
EXPLOSION HAZARD! The electrical components are not explosionproof. Do not use in an area where flammable
gases are present!
11.2. STarT-up/SelF TeST
1. Connect the pump to mains. Switch on device with power switch. LED STANDBY (4) will be on. Do not connect tubing set
yet! Do not connect footswitch yet!
2. Activate the device by pressing button ON / OFF (5). A short beep sounds.
3. The device enters self test.
Self test:
a. All LEDs and displays are turned on for 1 sec.
b. IRRIGATION displays (9) show program version for 1 sec. Then segments of left IRRIGATION display are turned on as
follows:
c. 1. Segment: test of footswitch sensor
d. 1.-2. Segment: test of pressure sensor IRRIGATION
e. 1.-3. Segment: test of pressure sensor SUCTION
f . 1.-4. Segment: test of pressure generator
g. 1.-5. test of vacuum generator
When self test is passed correctly a triple beep sounds. LEDs ON and STOP will be on and displays show preset pressure and
preset vacuum at last pump start.
4. Connect the tubing set (see chapter 9.4.). The valves of irrigation-suction instrument must be closed.
5. Perform functional check.
11.3. FunCTIonal TeST
The functional check must be carried out prior to each operation.
Irrigation:
1. Push button UP/DOWN IRRIGATION (8) to adjust preset pressure to maximum.
2. Start the pump.
3. Pressure increases to maximum within ca. 2 min. and the pressure generator is switched off. If preset pressure is not
obtained check tubes and connections to be tight.
4. Operate irrigation valve of instrument and fill the tubing set and the instrument completely with liquid until liquid
emerges from the instrument.
5. The functional check of IRRIGATION is complete.
Suction:
1. Push button UP/DOWN SUCTION (11) to adjust preset vacuum to maximum.
2. Vacuum increases to maximum within ca. 2 min. and the vacuum generator is switched off. If preset vacuum is not
obtained check tubes and connections to be tight.
3. The functional check of SUCTION is complete. Press the button START/STOP to stop the pump. The STOP LED lights
up. Now the device is tested and ready to be used in the operating room.
11.4. operaTIon oF THe DevICe
1 . After performing the functional check the pump can be started.
2. The pump is activated.
3. Pressure and vacuum generators are operating until preset values are obtained within the bottles. Values can be
adjusted during operation.
4. Operate valves on instrument to activate IRRIGATION or SUCTION according to your need. Pressure and vacuum
generators are switched off automatically if preset pressure and vacuum are obtained within the bottles.