9.4. ConneCTIng THe TubIng SeT
Tubing set has to be sterile, see chapter 9.2.
Step 1 - Tube 1, blue, irrigation:
Connect disposable sterile filter to female LL-connector of
spike. Connect male LLconnector of the tube to disposable
sterile filter. Connect female LL-connector of the tube to
connector IRRIGATION on pump.
Step 2- Tube 2, blue, irrigation:
Connect one side of the tube to tube connector on spike and
the other side of the tube to irrigation connector on instrument.
Step 3 - Tube 3, red, suction:
Connect disposable sterile filter to connector SUCTION on pump.
Connect male LLconnector of the tube to disposable sterile
filter. Connect the other side of the tube to connector „VACUUM“
on secretion container.
Step 4 - Tube 4, red, suction:
Connect one side of the tube to connector „PATIENT“ on secretion
container and the other side of the tube to suction connector on
instrument.
9.5. removIng THe TubIng SeT
First of all disconnect tubes from pump to release pressure
from the bottles. Separate all other connections then. Remove
and dispose sterile filters. Sterile filter is a disposable product
and not suitable For reprocessing. The manufacturer is released
from any product liability if products Intended for single use are
processed and re-sterilized contrary to these instructions.
Reprocessing of tubing set see chapter 9.2.