6.4. InSerTIg THe TubIng SeT
Tubing set has to be completely assembled and sterile, see chapter 6.3.
Step 1:
Take the blue irrigation tube. Place the tube stopper (blue) in the
upper tube holder (marked blue). Pull the pump segment taught
around the roller wheel and route it through the lower tube holder
until the second tube stopper (colorless) locks firmly into place.
Ensure the tube stoppers are correctly seated.
Step 2:
Take the red suction tube. Place the tube stopper (red) into the
upper tube holder (marked red). Proceed as described in step 1.
Step 3:
Next, connect all connectors on the suction and irrigation tubes
to the instrument, to the reservoir and to the fluid collection receptacle.
6.5. removIng THe TubIng SeT
Detach all connectors between the instrument and reservoir before
detaching the tube from the roller wheels.
Take the pump segment of the red suction tube out of the tube holder
first, and then the pump segment of the blue irrigation tube.
Recondition reusable tube sets for later use (see section 6.2.)
6.6. uSIng THe DevICe In operaTIng room - IlluST raTIon