DoUbLE RoLLER PUMP foR LAPARoScoPY
Ackermann
Page: 6 // 18
• Liability may be voided by misuse, abuse, violation of instructions for use, omission of required maintenance, opening
the equipment modifications, repairs by unauthorized persons! Use of other than original accesaccessories shall void
warranty and liability.
• Equipment has to be used at all times in accordance with the operating instructions!
This device complies with currently valid EMV standards. However tests of EMV standard DIN EN 60601-1-2:1993-05 are
carried out. Nevertheless, EMV interference may lead to malfunctioning. If malfunctioning is observed, ensure that the
device has been installed and is being operating in accordance with EMV guidelines for use supplied with the device. Note
that portable and mobile high-frequency communication equipment may influence the operation of this device. If the
device is used in the vicinity of or stacked together with other devices, the device or system must be monitored to ensure
standard use in accordance with these guidelines. If necessary, notify the manufacturer of the corresponding details.
IMPoRTANT!
In case of failure switch off the device at its main power switch (1 ). Pull mains connection.
Danger!
Do not open the unit. No user-servicable parts inside.
Symbol and description
Symbol and description
Symbol and description
Follow the instructions for use
Type BF applied part
Year of production
Legal manufacturer
mdc medical device certification GmbH,
Kriegerstraße 6,
70191 Stuttgart, Germany
Equipotential
(equalising potential)
Caution!
Hazardous voltage
Batch number
Article number
Do not use if package is
damaged
Spray protection (protection
against vertically falling drops)
Protective earthing
Product contains phthalate
6. HanDlIng of TubIng SeT
6.1. CompoSITIon of TubIng SeT
The tube set comprises two separate, color-coded tubes.
The blue tube is used for irrigation and the red one for suction. Both tubes are fitted with colored rings (blue and red).
You will find a colored ring at tube stopper of pump segment, see section 6.3.
To insert the correct tubes you will find corresponding color-coding on the device’s tube holder in blue and red.
6.2. reproCeSSIng of reuSable TubIng SeT
All components of the reusable tube set can be autoclaved at a maximum temperature of 134°C and pressure of 2 bar.
The tube set can be dismantled (see section 6.3.), but must be completely reassembled and dried before sterilization.
IMPoRTANT!
When fitting together the tube set, please ensure that there is a tight, leak-proof connection between the
adapter and the tube sections.